The soft x-ray Opacity Spectrometer (OpSpec) used on the National Ignition Facility (NIF) has recently incorporated an elliptically shaped crystal. The original OpSpec used two convex cylindrical crystals for time-integrated measurements of point-projection spectra from 540 to 2100 eV. However, with the convex geometry, the low-energy portion of the spectrum suffered from high backgrounds due to scattered x-rays as well as reflections from alternate crystal planes. An elliptically shaped crystal allows an acceptance aperture at the crossover focus between the crystal and the detector, which reduces background and eliminates nearly all reflections from alternate crystal planes. The current elliptical design is an improvement from the convex cylindrical design but has a usable energy range from 900 to 2100 eV. In addition, OpSpec is currently used on 18 NIF shots/year, in which both crystals are typically damaged beyond reuse, so efficient production of 36 crystals/year is required. Design efforts to improve the existing system focus on mounting reliability, reducing crystal strain to increase survivability between mounting and shot time, and extending the energy range of the instrument down to 520 eV. The elliptical design, results, and future options are presented.
1 November 2022 in Review of Scientific Instruments
The Opacity Platform on the National Ignition Facility (NIF) has been developed to measure opacities at varying densities and temperatures relevant to the solar interior and thermal cooling rates in white dwarf stars. The typical temperatures reached at NIF range between 150 and 210 eV, which allow these measurements to be performed experimentally. The captured opacities are crucial to validating radiation-hydrodynamic models that are used in astrophysics. The NIF opacity platform has a unique new capability that allows in situ measurement of the sample expansion. The sample expansion data are used to better understand the plasma conditions in our experiments by inferring the sample density throughout the duration of the laser drive. We present the details of the density measurement technique, data analysis, and recent results for Fe and MgO.
1 October 2022 in Review of Scientific Instruments
When compared with the National Ignition Facility’s (NIF) original soft x-ray opacity spectrometer, which used a convex cylindrical design, an elliptically shaped design has helped to increase the signal-to-noise ratio and eliminated nearly all reflections from alternate crystal planes. The success of the elliptical geometry in the opacity experiments has driven a new elliptical geometry crystal with a spectral range covering 520–1100 eV. When coupled with the primary elliptical geometry, which spans 1000–2100 eV, the new sub-keV elliptical geometry helps to cover the full iron L-shell and major oxygen transitions important to solar opacity experimentation. The new design has been built and tested by using a Henke x-ray source and shows the desired spectral coverage. Additional plans are underway to expand these opacity measurements into a mode of time-resolved detection, ∼1 ns gated, but considerations for the detector size and photometrics mean a crystal geometry redesign. The new low-energy geometry, including preliminary results from the NIF opacity experiments, is presented along with the expansion plans into a time-resolved platform.
Precise characterization of experimental radiation flow is required to validate the high energy density physics models, numerical methods, and codes that are used to simulate radiation-hydrodynamics phenomena such as thermal radiation transport in stochastic media. The Cassio code is used to simulate thermal radiation flow through inhomogeneous, stochastic-media-foam configurations containing optically thick clumps dispersed within an optically thin background aerogel. Cassio can model small inhomogeneous problems directly, but most problems require approximations to meet computer limitations on run-times and memory usage. Various examples of these approximations are methods that produce, in one calculation, an ensemble-averaged solution and associated standard deviation; reduced spatial dimensionality with approximate geometries; and full material homogenization with no geometric detail. Cassio simulations are used to design experiments at the OMEGA-60 Laser Facility that can measure the radiation flow using the spatially resolved COAX absorption spectroscopy diagnostic. The experimental platforms flow radiation through foam targets ranging from a background-only aerogel, to a single configuration of a specified stochastic medium, to a fully homogenized foam of the background and clump materials. Under constant total clump mass, larger clumps (here, larger than 10 μm diameter) will mix more slowly with the background such that the bulk radiation flow is faster than it would be in a fully homogenized material. The COAX platform can be used to infer temperature and density profiles in both the background material and clumps, simultaneously, and therefore to differentiate radiation flow in a range of stochastic and homogeneous media.
Radiation flow through an inhomogeneous medium is critical in a wide range of physics and astronomy applications, from transport across cloud layers on the Earth to the propagation of supernova blast waves producing ultraviolet and X-ray emission. Radiation interacts with matter, driving hydrodynamic feedback that further alters the radiation characteristics (energy and angular distribution). This paper reviews the current state of the art in the modeling of inhomogeneous radiation transport, subgrid models developed to capture this often-unresolved physics, and the experiments designed to improve our understanding of these models. This paper focuses on simulations based on upcoming experiments designed to test this physics. We present a series of detailed simulations (both single-clump and multi-clump conditions) probing the dependence on the physical properties of the radiation front (e.g., radiation energy) and material characteristics (specific heat, opacity, clump densities). We find that, unless the radiation pressure is high, the clumps will heat and then expand, effectively cutting off the radiation flow. The expanding winds can also produce shocks that generates high-energy emission. We compare our detailed simulations with some of the current subgrid prescriptions, identifying some of the limitations of these current models.
Predicting and modeling the behavior of experiments with radiation waves propagating through low-density foams require a detailed quantification of the numerous uncertainties present. In regimes where a prominent radiative shock is produced, key dynamical features include the shock position, temperature, and curvature and the spatial distribution and temperature of the corresponding supersonic radiation wave. The COAX experimental platform is designed to constrain numerical models of such a radiative shock propagating through a low-density foam by employing radiography for spatial and shock information, Dante for characterizing the x-ray flux from the indirectly driven target, and a novel spectral diagnostic designed to probe the temperature profile of the wave. In this work, we model COAX with parameterized 2D simulations and a Hohlraum-laser modeling package to study uncertainties in diagnosing the experiment. The inferred temperature profile of the COAX radiation transport experiments has been shown to differ from simulations more than expected from drive uncertainties that have been constrained by simultaneous soft x-ray flux and radiography measurements.