Brain–machine interface (BMI) provides an alternative route for controlling an external device with one’s intention. For individuals with motor-related disability, the BMI technologies can be used to replace or restore motor functions. Therefore, BMIs for movement restoration generally decode the neural activity from the motor-related brain regions. In this study, however, we designed a BMI system that uses sensory-related neural signals for BMI combined with electrical stimulation for reward. Four-channel electrocorticographic (ECoG) signals were recorded from the whisker-related somatosensory cortex of rats and converted to extract the BMI signals to control the one-dimensional movement of a dot on the screen. At the same time, we used operant conditioning with electrical stimulation on medial forebrain bundle (MFB), which provides a virtual reward to motivate the rat to move the dot towards the desired center region. The BMI task training was performed for 7 days with ECoG recording and MFB stimulation. Animals successfully learned to move the dot location to the desired position using S1BF neural activity. This study successfully demonstrated that it is feasible to utilize the neural signals from the whisker somatosensory cortex for BMI system. In addition, the MFB electrical stimulation is effective for rats to learn the behavioral task for BMI.
Objective. Connectors for implantable neural prosthetic systems provide several advantages such as simplification of surgery, safe replacement of implanted devices, and modular design of the implant systems. With the rapid advancement of technologies for neural implants, miniaturized multichannel implantable connectors are also required. In this study, we propose a reconnectable and area-efficient multichannel implantable connector. Approach. A female-to-female adapter was fabricated using the thermal-press bonding of micropatterned liquid crystal polymer films. A bump inside the adapter enabled a reliable electrical connection by increasing the contact pressure between the contact pads of the adapter and the inserted cable. After connection, the adapter is enclosed in a metal case sealed with silicone elastomer packing. With different sizes of the packings, leakage current tests were performed under accelerated conditions to determine the optimal design for long-term reliability. Repeated connection tests were performed to verify the durability and reconnectability of the fabricated connector. The connector was implanted in rats, and the leakage currents were monitored to evaluate the stability of the connector in vivo. Main results. The fabricated four- and eight-channel implantable connectors, assembled with the metal cases, had a diameter and length of 6 and 17 mm, respectively. Further, the contact resistances of the four- and eight-channel connectors were 53.2 and 75.2 mΩ, respectively. The electrical contact remained stable during repeated connection tests (50 times). The fabricated connectors with packings having 125%, 137%, and 150% volume ratios to the internal space of the metal case failed after 14, 88, and 14 d, respectively, in a 75 °C saline environment. In animal tests with rats, the connector maintained low leakage current levels for up to 92 d. Significance. An implantable and reconnectable multichannel connector was developed and evaluated. The feasibility of the proposed connector was evaluated in terms of electrical and mechanical characteristics as well as sealing performance. The proposed connector is expected to have potential applications in implantable neural prosthetic systems.