13 February 2024 in Composites Part B: Engineering
2 December 2022 in IEEE Sensors Journal
This work focuses on the manufacturing and testing of a new device for medical bandages monitoring. An excessive pressure exerted from the compression bandage can block the blood flow of the patient, causing different medical complications to skin, nerves and to the circulatory system. On the contrary, if the pressure applied is low, the therapy is not effective. The utility therefore arises of a device capable of quantitatively indicating the correct adjustment of the bandage. The technological demonstrators developed consist of a polyurethane elastomeric shell with a thin composite supporting core. Fiber Bragg Grating Sensors (FBGS) embedded within this core permit the detection of the sub-bandage pressure applied during compression therapy. The two prototypes were applied under arm bandages to evaluate their capability to transmit the applied pressure to the embedded FBGS. We demonstrated the ability to monitor the bandaging action measuring the level of pressure exerted with the rounds of bandages. Moreover, the thin membranes permit the monitoring of the heartbeat of the patient, giving feedback about blood irroration. The device developed is therefore promising to improve the results of compression therapy.