1 October 2005 in The Journal of the Acoustical Society of America
A time-domain finite-difference solution to the equations of linear elasticity was used to model the propagation of lithotripsy waves in kidney stones. The model was used to determine the loading on the stone (principal stresses and strains and maximum shear stresses and strains) due to the impact of lithotripsy shock waves. The simulations show that the peak loading induced in kidney stones is generated by constructive interference from shear waves launched from the outer edge of the stone with other waves in the stone. Notably the shear wave induced loads were significantly larger than the loads generated by the classic Hopkinson or spall effect. For simulations where the diameter of the focal spot of the lithotripter was smaller than that of the stone the loading decreased by more than 50%. The constructive interference was also sensitive to shock rise time and it was found that the peak tensile stress reduced by 30% as rise time increased from . These results demonstrate that shear waves likely play a critical role in stone comminution and that lithotripters with large focal widths and short rise times should be effective at generating high stresses inside kidney stones.
1 August 2007 in Ultrasound in Medicine & Biology
1 September 2004 in Journal of Endourology
Background and Purpose: During SWL, stone motion secondary to ventilatory motion can be as much as 50 mm. This is much larger than the 4- to 15-mm diameter of the focal regions on most clinical machines. The goal of this study was to determine the effect of stone motion on the fragmentation efficiency of a clinical lithotripter with a small (4-mm) focal spot. Materials and Methods: A model stone (6.5 × 7.5 mm) made of gypsum cement was used as an in vitro target for a Storz Modulith SLX lithotripter with a custom-designed waterbath. A motorized positioner was used to translate the stone in order to simulate ventilatory motion. The excursion was variable up to 48 mm (± 24 mm about the focus). After treatment by 400 shockwaves, the remnants (>2 mm) were dried and weighed. Results: Fragmentation efficiency was reduced (P20 mm, it appeared that three-quarters of the shockwaves missed the stone. These data imply that ventilatory gating or stone tracking may result in fewer shockwaves being required for successful treatment with this lithotripter.
1 February 1998 in Ultrasound in Medicine & Biology
Stone comminution and tissue damage in lithotripsy are sensitive to the acoustic field within the kidney, yet knowledge of shock waves in vivo is limited. We have made measurements of lithotripsy shock waves inside pigs with small hydrophones constructed of a 25-microm PVDF membrane stretched over a 21-mm diameter ring. A thin layer of silicone rubber was used to isolate the membrane electrically from pig fluid. A hydrophone was positioned around the pig kidney following a flank incision. Hydrophones were placed on either the anterior (shock wave entrance) or the posterior (shock wave exit) surface of the left kidney. Fluoroscopic imaging was used to orient the hydrophone perpendicular to the shock wave. For each pig, the voltage settings (12-24 kV) and the position of the shock wave focus within the kidney were varied. Waveforms measured within the pig had a shape very similar to those measured in water, but the peak pressure was about 70% of that in water. The focal region in vivo was 82 mm x 20 mm, larger than that measured in vitro (57 mm x 12 mm). It appeared that a combination of nonlinear effects and inhomogeneities in the tissue broadened the focus of the lithotripter. The shock rise time was on the order of 100 ns, substantially more than the rise time measured in water, and was attributed to higher absorption in tissue.
1 September 2007 in The Journal of the Acoustical Society of America
A three-dimensional model of the forward propagation of nonlinear sound beams in inhomogeneous media, a generalized Khokhlov-Zabolotskaya-Kuznetsov equation, is described. The Texas time-domain code (which accounts for paraxial diffraction, nonlinearity, thermoviscous absorption, and absorption and dispersion associated with multiple relaxation processes) was extended to solve for the propagation of nonlinear beams for the case where all medium properties vary in space. The code was validated with measurements of the nonlinear acoustic field generated by a phased array transducer operating at 2.5MHz in water. A nonuniform layer of gel was employed to create an inhomogeneous medium. There was good agreement between the code and measurements in capturing the shift in the pressure distribution of both the fundamental and second harmonic due to the gel layer. The results indicate that the numerical tool described here is appropriate for propagation of nonlinear sound beams through weakly inhomogeneous media.
28 February 2011 in Ultrasound in Medicine & Biology