Shusuke Yoshimoto

Shusuke Yoshimoto

Auto-generated profile, based on publicly available information

Shusuke Yoshimoto

SANKEN (The Institute of Scientific and Industrial Research), Osaka University, Kobe, Japan

Research Focus & Metrics

Academic Career Length

18Years (2009-Present)

(Calculated once Scilit publication data is synced: current year − year of first publication + 1)

Representative Publications

View All Publications
26 January 2022 in IEEE Access
Verified
Journal Article
Open access
SMShoya Matsumori
Shoya Matsumori
KTKosei Teramoto
Kosei Teramoto
HIHiroya Iyori
Hiroya Iyori
TSTakanori Soda
Takanori Soda
SYShusuke Yoshimoto
Shusuke Yoshimoto
HMHaruo Mizutani
Haruo Mizutani
Analysis of sleep stages using electroencephalograms (EEGs), a critical procedure in health monitoring, has been researched extensively. Scoring of the sleep stages is highly dependent on experts’ knowledge or hand-crafted features created by experts. Recent advancements in deep learning have, nevertheless, successfully automated the scoring process with accuracy comparable to human performance. Previous methods were less convenient as they employed specific EEG devices designed for the experiments. Such devices are usually expensive and require a special environment and arrangements for their use. In this study, we propose an automatic sleep-stage scoring system using a patch-type wearable EEG sensor based on deep learning. The proposed sensor, namely Haru, is lightweight, inexpensive, and easy to use. As for the training architecture, we implement a model based on the state-of-the-art deep learning algorithm specifically designed for automatic sleep staging. We record three channels of sleep EEGs from 30 subjects at a sampling rate of 250 Hz with the proposed system and a polysomnogram (PSG) device simultaneously. Training datasets are created on the basis of the annotated samples in PSG measurement. In the experiment, we train our model and evaluate the results using the leave-one-out cross-validation method. The results indicate that the proposed Haru Sleep is capable of scoring sleep stages with a scoring accuracy of 78.6% and an F1 score of 73.4%, which is comparable to that obtained by clinical PSG devices.

1 September 2010
Verified
Conference Paper
Closed access
SOShunsuke Okumura
Shunsuke Okumura
SYShusuke Yoshimoto
Shusuke Yoshimoto
KYKosuke Yamaguchi
Kosuke Yamaguchi
YNYohei Nakata
Yohei Nakata
HKHiroshi Kawaguchi
Hiroshi Kawaguchi
MYMasahiko Yoshimoto
Masahiko Yoshimoto
This paper proposes 7T SRAM which realizes block-level simultaneous copying feature. The proposed SRAM can be used for data transfer between local memories such as checkpoint data storage and transactional memory. The 1-Mb SRAM is comprised of 32-kb blocks, in which 16-kb data can be copied in 33.3 ns at 1.2V. The proposed scheme reduces energy consumption in copying by 92.7% compared to the conventional read-modify-write manner.

1 March 2009
Verified
Conference Paper
Closed access
SOShunsuke Okumura
Shunsuke Okumura
YIYusuke Iguchi
Yusuke Iguchi
SYShusuke Yoshimoto
Shusuke Yoshimoto
HFHidehiro Fujiwara
Hidehiro Fujiwara
HNHiroki Noguchi
Hiroki Noguchi
MYMasahiko Yoshimoto
Masahiko Yoshimoto
We present a small-area 10T SRAM cell without half selection problem. As well, the proposed 10T cell achieves a faster access time and low voltage operation. The cell area is reduced by 25%, and the cell current is increased by 21%, compared with the prior 10T cell. The minimum operating voltage is lowered by the column line assist (CLA) scheme that suppresses write margin degradation. By measurement, we confirmed that the proposed 128-kb SRAM works at 0.56 V.

1 March 2022 in Bioelectricity
Verified
Journal Article
Closed access
FLFangzhou Li
Fangzhou Li
SMShoya Matsumori
Shoya Matsumori
NENaohiro Egawa
Naohiro Egawa
SYShusuke Yoshimoto
Shusuke Yoshimoto
KYKotaro Yamashiro
Kotaro Yamashiro
RTRyosuke Takahashi
Ryosuke Takahashi
Background: Developing a screening method for mild cognitive impairment in the aging population and intervening early in the progression of dementia based on such a method, remains challenging. Electroencephalography (EEG) is a noninvasive and sensitive tool to assess the functional activity of the brain, and wireless and mobile EEG (wmEEG) could serve as an alternative screening technique that is widely tolerable in patients with dementia from the preclinical to severe stage. Materials and Methods: Using wmEEG, we recorded bioelectrical activity (BA) from the forehead in 101 individuals with dementia and nondementia controls (NCs) during 4 tasks and investigated which task could differentiate dementia from NC. Results: We found significant differences in three power spectra of the time–frequency analysis (3–4, 5–7, and 17–23 Hz) between dementia and NC under an eyes-open condition and a significant consistent difference in a specific slow alpha power spectrum (6–8 Hz) between Alzheimer's disease (AD) and dementia with Lewy bodies (DLB) under an eyes-closed condition. These results were confirmed by classification analysis using a deep learning method based on the whole wmEEG data sets, in which the accuracy of discriminating dementia from NC under the eyes-open condition was higher than that under the eyes-closed condition (0.71 vs. 0.52, respectively). Moreover, the accuracy of discriminating AD from DLB under the eyes-closed condition was higher than that under the eyes-open condition (0.77 vs. 0.64, respectively). Conclusion: The result of this pilot study suggests that wmEEG can be a useful tool for recording BA, and that analyzing BA may help to detect early dementia and discriminate dementia subtypes effectively and objectively.

24 August 2023 in IEEE Sensors Letters
Verified
Journal Article
Closed access
MIMasaki Ishibashi
Masaki Ishibashi
SIShintaro Izumi
Shintaro Izumi
RTRyo Takamatsu
Ryo Takamatsu
SYShusuke Yoshimoto
Shusuke Yoshimoto
YNYuki Noda
Yuki Noda
HKHiroshi Kawaguchi
Hiroshi Kawaguchi
The use of piezoelectric film for pulse-wave measurement is advantageous in terms of power consumption, but it has limitations in terms of sensitivity and noise resistance. This study described a high-sensitivity pulse-wave sensor that used a piezoelectric film. Furthermore, we proposed a three-dimensional gel structure to improve the sensor's skin contact condition and measurement sensitivity through structural innovations. Our approach was to increase the strain of the piezoelectric film by utilizing a structure with a tapered rate, which increased the sensitivity of the sensor structure. The subject was a healthy adult male (24 years old). We also fabricated a prototype pulse-wave sensor with the proposed structure and continuously measured pulse waves at multiple locations on the radial artery of the wrist. Through signal processing, we were able to obtain measurement data that were effective for monitoring pulse wave propagation and showed that fluctuations in pulse wave velocity could be detected.