This paper presents the validation of the design of a 6-in. dual-wavelength transmission sphere (TS) based on the Fizeau interferometer. The TS was verified at a wavelength of 632.8 nm and is compatible for measuring the transmitted wavefront error of an i-line lithography lens. The achromatic design is imperative for a dual wavelength TS and requires more lenses for correcting the chromatic focal shift. Because the overall weight of the TS should be controlled within the load range of the piezoelectric transducers of the interferometer, the fabrication and mounting of lenses with high aspect ratios are challengeable. The mounting of the reference surface based on three-point mounting was successful for the residual reference wavefront error under peak-to-valley (PV) . Furthermore, the reference wavefront is typically restricted within PV to avoid distorted interference fringes. Therefore, we built a double-pass interferometer model for tolerance analysis, and the error budget facilitated decision-making regarding the suitable specifications of lens manufacturing and assembly for cost efficiency. The test results demonstrated that the deformation of the reference wavefront and the residual reference wavefront error met the critical specification at 632.8 nm and that the achromatic TS is compatible for measuring i-line lithography lenses.
An absolute measurement method involving a computer-generated hologram to facilitate the identification of manufacturing form errors and mounting- and gravity-induced deformations of a 300-mm aspheric mirror is proposed. In this method, the frequency and magnitude of the curve graph plotted from each Zernike coefficient obtained by rotating the mirror with various orientations about optical axis were adopted to distinguish the nonrotationally symmetric aberration. In addition, the random ball test was used to calibrate the rotationally symmetric aberration (spherical aberration). The measured absolute surface figure revealed that a highly accurate aspheric surface with a peak-to-valley value of 1/8 wave at 632.8 nm was realized after the surface figure was corrected using the reconstructed error map.