The interaction between light and biological tissue is determined by the intensity and wavelength of the light and by the optical properties (absorption and scattering) of the tissue. Biological tissues can be penetrated relatively deeply by near-infrared (NIR) light, and thus the brain can be studied. In the present work NIR-LEDs (light-emitting diodes) are used to couple light into the head tissue. The light is strongly scattered, because the brain is a strong scatterer in the NIR.
1, 2 The intensity of the re-emerging light is measured by a photodiode detector, and thus the optical properties of the brain can be obtained. Every tissue also has its own absorption spectrum. Biological tissues contain three main absorbers: water, hemoglobin, and lipid. Below a wavelength of 650 nm, strong absorption occurs for oxyhemoglobin (O
2Hb) and deoxyhemoglobin (HHb), whereas water shows strong absorption above 950 nm. As a consequence of this, light does not penetrate deeply into the tissue in these regions. To be able to interrogate deep layers of tissue, near-infrared spectroscopy (NIRS) uses NIR wavelengths between 650 and 950 nm. NIRS is primarily employed to measure tissue oxygenation with parameters such as O
2Hb, HHb, total hemoglobin (tHb) concentration, tissue oxygen saturation (StO
2), and cerebral blood flow.
3 NIRS has the advantage to enable bedside continuous and noninvasive monitoring of these parameters. Furthermore, it can be combined with other types of cerebral monitoring such as EEG or aEEG, and it is cost effective.
2