9 July 2026 in SSRN Electronic Journal
Ti and its alloys are standard materials providing a bioinert surface for osseointegrated implant devices. However, developing bioactive surfaces for such applications is a great challenge in biomaterial surface engineering. This study investigates the surface modification of titanium by depositing thin films of niobium and titanium-niobium alloys using magnetron sputtering. The overall objective is to evaluate the effect of Nb introduction on the physicochemical, structural, and biological properties of the surface. Analysis via scanning electron microscopy and atomic force microscopy demonstrated that thermal treatment at moderate temperatures (500–750°C) promotes grain growth and alters surface roughness. X-ray diffraction indicated the coexistence of phases and temperature-dependent structural alterations. Contact angle measurements confirmed the surface hydrophilicity, with values suitable for biomedical applications. In vitro assays with MC3T3-E1 pre-osteoblastic cells demonstrated biocompatibility with cell viability >70%. Furthermore, alkaline phosphatase activity and extracellular matrix mineralization were satisfactory over 14 and 21 days. Although the surfaces did not show direct antimicrobial activity against Staphylococcus aureus, this neutrality is considered advantageous for preserving the peri-implant biological ecosystem. Therefore, the developed Nb and Ti-Nb coatings prove to be a promising and innovative strategy for enhancing the bioactivity of metallic implants.