18 May 2021 in IEEE Transactions on Nuclear Science
We present the design, fabrication and radiation testing of a custom 20-channel application specific integrated circuit (ASIC) for amplification and digitization of microvolt signal levels from space instrumentation (e.g., thermopile sensors) operating in harsh space radiation environments. The ASIC is fabricated in a commercial 180 nm CMOS process node using a combination of radiation hard by design and discrete-time analog signal processing to mitigate leakage, offset, 1/f and thermal noise. The chip has an input referred integrated noise voltage of 190 nV in a 240 Hz bandwidth. Irradiation with very high total ionizing dose and high-energy heavy ions was carried out to evaluate the chip’s radiation hardness. The chip’s post radiation performance evidenced parametric immunity to greater than 50 Mrad (Si) total ionizing dose and no single event latch-up up to at least 123.6MeV-cm2/mg linear energy transfer.
12 January 2026 in IEEE Transactions on Applied Superconductivity
We present the design and testing of a prototype multiplexing kinetic inductance detector (KID) readout electronics for the PRobe far-Infrared Mission for Astrophysics (PRIMA) space mission. PRIMA is a Probe-class astrophysics mission concept that will answer fundamental questions about the formation of planetary systems, the co-evolution of stars and supermassive black holes in galaxies, and the rise of heavy elements and dust over cosmic time. The readout electronics for PRIMA must be compatible with operation at Earth-Sun L2 and capable of multiplexing more than 1000 detectors over 2.5 GHz bandwidth while consuming around 30 W per readout chain. The electronics must also be capable of switching between the two instruments, which have different readout bands: the hyperspectral imager (PRIMAger, 2.6-4.9 GHz) and the spectrometer (FIRESS, 0.4-2.4 GHz). The PRIMA readout electronics use high-heritage SpaceCube digital electronics with a build-to-print SpaceCube Mini v3.0 board using a radiation-tolerant Kintex KU060 field programmable gate array (FPGA) and a custom high-speed digitizer board, along with RF electronics that provide filtering and power conditioning. We present the driving requirements for the system, as well as the hardware, firmware, software, and system-level design that meets those requirements.
23 May 2022 in IEEE Sensors Journal
A Non-Dispersive Infra-Red (NDIR) gas analyzer has been designed, and a proof-of concept built and validated, to continuously monitor water vapor and carbon dioxide gas partial pressures, for a space-based instrument application, that requires the safe transfer of sublimated volatiles from a comet sample containment system to a gas containment system within the operational pressure-temperature conditions, e .g., water vapor, in the range 3 Torr (268 K) and 30 mTorr (223 K). The NDIR gas analyzer uses quad thermopile detectors integrated with spectral bandpass comprising filters centered at $2.7 ~ \mu \text{m}$ and $6.58 ~ \mu \text{m}$ for water absorption channels, spectral bandpass filter centered at $4.25 ~ \mu \text{m}$ for carbon dioxide absorption channel, and spectral band-pass filter centered at $4.0 ~ \mu \text{m}$ for the reference channel, paired with a pulsed infra-red source. The thermopile detector signals are measured and processed using a custom-built radiation-hard-by-design multi-channel digitizer (MCD) application specific integrated circuit (ASIC). The NDIR gas analyzer can detect water vapor partial pressures up to 3 Torr with a 30 mTorr resolution (1 sigma) and carbon dioxide partial pressures up to 175 Torr with a 50 mTorr resolution (1 sigma) within the temperature range 223–268 K.