To evaluate prostate intrafraction motion using MRI during the full course of online adaptive MR-Linac radiotherapy (RT) fractions, in preparation of MR-guided extremely hypofractionated RT. Five low and intermediate risk prostate cancer patients were treated with 20 x 3.1 Gy fractions on a 1.5T MR-Linac. Each fraction, initial MRI (Pre) scans were obtained at the start of every treatment session. Pre-treatment planning MRI contours were propagated and adapted to this Pre scan after which plan re-optimization was started in the treatment planning system followed by dose delivery. 3D Cine-MR imaging was started simultaneously with beam-on and acquired over the full beam-on period. Prostate intrafraction motion in this cine-MR was determined with a previously validated soft-tissue contrast based tracking algorithm. In addition, absolute accuracy of the method was determined using a 4D phantom. Prostate motion was completely automatically determined over the full on-couch period (approx. 45 min) with no identified mis-registrations. The translation 95% confidence intervals are within clinically applied margins of 5 mm, and plan adaption for intrafraction motion was required in only 4 out of 100 fractions. This is the first study to investigate prostate intrafraction motions during entire MR-guided RT sessions on an MR-Linac. We have shown that high quality 3D cine-MR imaging and prostate tracking during RT is feasible with beam-on. The clinically applied margins of 5 mm have proven to be sufficient for these treatments and may potentially be further reduced using intrafraction plan adaptation guided by cine-MR imaging.
1 July 2020 in Physics and Imaging in Radiation Oncology
Monitoring the intrafraction motion and its impact on the planned dose distribution is of crucial importance in radiotherapy. In this work we quantify the delivered dose for the first prostate patients treated on a combined 1.5T Magnetic Resonance Imaging (MRI) and linear accelerator system in our clinic based on online 3D cine-MR and treatment log files. A prostate intrafraction motion trace was obtained with a soft-tissue based rigid registration method with six degrees of freedom from 3D cine-MR dynamics with a temporal resolution of 8.5–16.9 s. For each fraction, all dynamics were also registered to the daily MR image used during the online treatment planning, enabling the mapping to this reference point. Moreover, each fraction’s treatment log file was used to extract the timestamped machine parameters during delivery and assign it to the appropriate dynamic volume. These partial plans to dynamic volume combinations were calculated and summed to yield the delivered fraction dose. The planned and delivered dose distributions were compared among all patients for a total of 100 fractions. The clinical target volume underwent on average a decrease of 2.2% ± 2.9% in terms of D99% coverage while bladder V62Gy was increased by 1.6% ± 2.3% and rectum V62Gy decreased by 0.2% ± 2.2%. The first MR-linac dose reconstruction results based on prostate tracking from intrafraction 3D cine-MR and treatment log files are presented. Such a pipeline is essential for online adaptation especially as we progress to MRI-guided extremely hypofractionated treatments.
1 March 2021 in Clinical and Translational Radiation Oncology
Magnetic resonance (MR)-guided linear accelerator (MR-Linac) systems have changed radiotherapy workflows. The addition of daily online contour adaptation allows for higher precision treatment, but also increases the workload of those involved. We train radiation therapists (RTTs) to perform daily online contour adaptation for MR-Linac treatment of prostate cancer (PCa) patients. The purpose of this study was to evaluate these prostate contours by performing an interfraction and interobserver analysis. Clinical target volume (CTV) contours generated online by RTTs from 30 low-intermediate risk PCa patients, treated with 5x7.25 Gy, were used. Two physicians (Observers) judged the RTTs contours and performed adaptations when necessary. Interfraction relative volume differences between the first and the subsequent fractions were calculated for the RTTs, Observer 1, and Observer 2. Additionally, interobserver dice’s similarity coefficient (DSC) for fraction 2–5 was calculated with the RTTs- and physician-adapted contours. Clinical acceptability of the RTTs contours was judged by a third observer. Mean (SD) online contour adaptation time was 12.6 (±3.8) minutes and overall median (interquartile range [IQR]) relative volume difference was 9.3% (4.4–13.0). Adaptations by the observers were mostly performed at the apex and base of the prostate. Median (IQR) interobserver DSC between RTTs and Observer 1, RTTs and Observer 2, and Observer 1 and 2 was 0.99 (0.98–1.00), 1.00 (0.98–1.00), and 1.00 (0.99–1.00), respectively. Contours were acceptable for clinical use in 113 (94.2%) fractions. Dose-volume histogram (DVH) analysis showed significant CTV underdosage for one of the seven identified outliers. Daily online contour adaptation by RTTs is clinically feasible for MR-Linac treatment of PCa.
Intrafraction motion during radiotherapy limits margin reduction and dose escalation. Magnetic resonance (MR)-guided linear accelerators (MR-Linac) have emphasised this issue by enabling intrafraction imaging. We present and clinically apply a new workflow to counteract systematic intrafraction motion during MR-guided stereotactic body radiotherapy (SBRT). With the sub-fractionation workflow, the daily dose is delivered in multiple sequential parts (sub-fractions), each adapted to the latest anatomy. As each sub-fractionation treatment plan complies with the dose constraints, no online dose accumulation is required. Imaging and treatment planning are executed in parallel with dose delivery to minimise dead time, enabling an efficient workflow. The workflow was implemented on a 1.5 T MR-Linac and applied in 15 prostate cancer (PCa) patients treated with 5 × 7.25 Gy in two sub-fractions of 3.625 Gy (10 × 3.625 Gy in total). Intrafraction clinical target volume (CTV) motion was determined and compared to a workflow with single-plan delivery. Furthermore, required planning target volume (PTV) margins were determined. Average on-table time was 42.7 min. Except for two fractions, all fractions were delivered within 60 min. Average intrafraction 3D CTV displacement (±standard deviation) was 1.1 mm (± 0.7) with the sub-fractionation workflow, whereas this was up to 3.5 mm (± 2.4) without sub-fractionation. Calculated PTV margins required with sub-fractionation were 1.0 mm (left-right), 2.4 mm (cranial-caudal), and 2.6 mm (anterior-posterior). Feasibility of the sub-fractionation workflow was demonstrated in 15 PCa patients treated with two sub-fractions on a 1.5 T MR-Linac. The workflow allows for significant PTV margin reduction in these patients by reducing systematic intrafraction motion during SBRT.
Magnetic resonance (MR)-guided linear accelerators (MR-Linac) enable accurate estimation of delivered doses through dose accumulation using daily MR images and treatment plans. We aimed to assess the association between the accumulated bladder (wall) dose and patient-reported acute urinary toxicity in prostate cancer (PCa) patients treated with stereotactic body radiation therapy (SBRT). One-hundred-and-thirty PCa patients treated on a 1.5 T MR-Linac were included. Patients filled out International Prostate Symptom Scores (IPSS) questionnaires at baseline, 1 month, and 3 months post-treatment. Deformable image registration-based dose accumulation was performed to reconstruct the delivered dose. Dose parameters for both bladder and bladder wall were correlated with a clinically relevant increase in IPSS (≥ 10 points) and/or start of alpha-blockers within 3 months using logistic regression. Thirty-nine patients (30%) experienced a clinically relevant IPSS increase and/or started with alpha-blockers. Bladder D5cm3, V10-35Gy (in %), and Dmean and Bladder wall V10-35Gy (cm3 and %) and Dmean were correlated with the outcome (odds ratios 1.04-1.33, p-values 0.001-0.044). Corrected for baseline characteristics, bladder V10-35Gy (in %) and Dmean and bladder wall V10-35Gy (cm3 and %) and Dmean were still correlated with the outcome (odds ratios 1.04-1.30, p-values 0.001-0.028). Bladder wall parameters generally showed larger AUC values. This is the first study to assess the correlation between accumulated bladder wall dose and patient-reported urinary toxicity in PCa patients treated with MR-guided SBRT. The dose to the bladder wall is a promising parameter for prediction of patient-reported urinary toxicity and therefore warrants prospective validation and consideration in treatment planning.
Purpose: To describe the development and first outcomes of the Utrecht Prostate Cohort (UPC): the first ‘trials within cohorts’ (TwiCs) platform for prostate cancer (PCa). Methods: All non-metastasized, histologically proven PCa patients who are planned to receive standard of care are eligible for inclusion in UPC. Patients provide informed consent for the collection of clinical and technical patient data, physician-reported outcomes, and patient-reported outcomes (PROs) up to 10 years post-treatment. Additionally, patients may provide broad consent for future randomization for experimental-intervention trials (TwiCs). Changes in PROs (EPIC-26 questionnaire domains) of the participants who received standard of care were analyzed using Wilcoxon signed-rank tests. Results: In two years, 626 patients were enrolled, 503 (80.4%) of whom provided broad consent for future randomization. Among these, 293 (46.8%) patients underwent magnetic resonance-guided adaptive radiotherapy (MRgRT), 116 (18.5%) CT-guided external beam radiation therapy (EBRT), 109 (17.4%) robot-assisted radical prostatectomy (RARP), and 65 (10.4%) patients opted for active surveillance. Patients treated with MRgRT and CT-guided EBRT showed a transient but significant decline in urinary irritative/obstructive and bowel domain scores at 1-month follow-up. RARP patients showed a significant deterioration of urinary incontinence domain scores between baseline and all follow-up moments and significant improvement of urinary irritative/obstructive domain scores between baseline and 9- and 12-month follow-up. All radical treatment groups showed a significant decline in sexual domain scores during follow-up. Active surveillance patients showed no significant deterioration over time in all domains. Conclusion: The first results from the UPC study show distinct differences in PROs between treatment options for PCa. Registration No.: NCT04228211.