Jody Christopher May

Jody Christopher May

Auto-generated profile, based on publicly available information

Jody Christopher May

Vanderbilt University

Research Focus & Metrics

Academic Career Length

24Years (2003-Present)

(Calculated once Scilit publication data is synced: current year − year of first publication + 1)

Representative Publications

View All Publications
1 August 2017 in Analytical Chemistry
Verified
Journal Article
Closed access
SSSarah Stow
Sarah Stow
TCTim J. Causon
Tim J. Causon
XZXueyun Zheng
Xueyun Zheng
RKRuwan T. Kurulugama
Ruwan T. Kurulugama
TMTeresa Mairinger
Teresa Mairinger
JFJohn C. Fjeldsted
John C. Fjeldsted

19 December 2014 in Analytical Chemistry
Verified
Journal Article
Open access
JMJody C. May
Jody C. May
JMJohn McLean
John McLean

22 January 2014 in Analytical Chemistry
Verified
Journal Article
Open access
JMJody C. May
Jody C. May
CGCody R. Goodwin
Cody R. Goodwin
NLNichole M. Lareau
Nichole M. Lareau
KLKatrina L. Leaptrot
Katrina L. Leaptrot
CMCaleb B. Morris
Caleb B. Morris
JMJohn McLean
John McLean
Ion mobility-mass spectrometry measurements which describe the gas-phase scaling of molecular size and mass are of both fundamental and pragmatic utility. Fundamentally, such measurements expand our understanding of intrinsic intramolecular folding forces in the absence of solvent. Practically, reproducible transport properties, such as gas-phase collision cross-section (CCS), are analytically useful metrics for identification and characterization purposes. Here, we report 594 CCS values obtained in nitrogen drift gas on an electrostatic drift tube ion mobility-mass spectrometry (IM-MS) instrument. The instrument platform is a newly developed prototype incorporating a uniform-field drift tube bracketed by electrodynamic ion funnels and coupled to a high resolution quadrupole time-of-flight mass spectrometer. The CCS values reported here are of high experimental precision (±0.5% or better) and represent four chemically distinct classes of molecules (quaternary ammonium salts, lipids, peptides, and carbohydrates), which enables structural comparisons to be made between molecules of different chemical compositions for the rapid “omni-omic” characterization of complex biological samples. Comparisons made between helium and nitrogen-derived CCS measurements demonstrate that nitrogen CCS values are systematically larger than helium values; however, general separation trends between chemical classes are retained regardless of the drift gas. These results underscore that, for the highest CCS accuracy, care must be exercised when utilizing helium-derived CCS values to calibrate measurements obtained in nitrogen, as is the common practice in the field.

27 November 2018 in Chemical Science
Verified
Journal Article
Open access
PDF file
JPJaqueline A. Picache
Jaqueline A. Picache
BRBailey S. Rose
Bailey S. Rose
ABAndrzej Balinski
Andrzej Balinski
KLKatrina L. Leaptrot
Katrina L. Leaptrot
SSStacy D. Sherrod
Stacy D. Sherrod
JMJohn McLean
John McLean
The Unified Compendium is an online interactive tool that utilizes ion mobility collision cross sections to annotate biochemical molecules.

1 November 2011 in Biochimica et Biophysica Acta (BBA) - Molecular and Cell Biology of Lipids
Verified
Journal Article
Closed access
MKMichal Kliman
Michal Kliman
JMJody C. May
Jody C. May
JMJohn McLean
John McLean

18 June 2016 in Analytical Chemistry
Verified
Journal Article
Open access
KHKelly Marie Hines
Kelly Marie Hines
JMJody C. May
Jody C. May
JMJohn A. McLean
John A. McLean
LXLibin Xu
Libin Xu
Collision cross section (CCS) measurement of lipids using traveling wave ion mobility-mass spectrometry (TWIM-MS) is of high interest to the lipidomics field. However, currently available calibrants for CCS measurement using TWIM are predominantly peptides that display quite different physical properties and gas-phase conformations from lipids, which could lead to large CCS calibration errors for lipids. Here we report the direct CCS measurement of a series of phosphatidylcholines (PCs) and phosphatidylethanolamines (PEs) in nitrogen using a drift tube ion mobility (DTIM) instrument and an evaluation of the accuracy and reproducibility of PCs and PEs as CCS calibrants for phospholipids against different classes of calibrants, including polyalanine (PolyAla), tetraalkylammonium salts (TAA), and hexakis(fluoroalkoxy)phosphazines (HFAP), in both positive and negative modes in TWIM-MS analysis. We demonstrate that structurally mismatched calibrants lead to larger errors in calibrated CCS values while the structurally matched calibrants, PCs and PEs, gave highly accurate and reproducible CCS values at different traveling wave parameters. Using the lipid calibrants, the majority of the CCS values of several classes of phospholipids measured by TWIM are within 2% error of the CCS values measured by DTIM. The development of phospholipid CCS calibrants will enable high-accuracy structural studies of lipids and add an additional level of validation in the assignment of identifications in untargeted lipidomics experiments.