The objective of the Third International Intercomparison on EPR Tooth Dosimetry was to evaluate laboratories performing tooth enamel dosimetry <300 mGy. Final analysis of results included a correlation analysis between features of laboratory dose reconstruction protocols and dosimetry performance. Applicability of electron paramagnetic resonance (EPR) tooth dosimetry at low dose was shown at two applied dose levels of 79 and 176 mGy. Most (9 of 12) laboratories reported the dose to be within 50 mGy of the delivered dose of 79 mGy, and 10 of 12 laboratories reported the dose to be within 100 mGy of the delivered dose of 176 mGy. At the high-dose tested (704 mGy) agreement within 25% of the delivered dose was found in 10 laboratories. Features of EPR dose reconstruction protocols that affect dosimetry performance were found to be magnetic field modulation amplitude in EPR spectrum recording, EPR signal model in spectrum deconvolution and duration of latency period for tooth enamel samples after preparation.
The use of electron paramagnetic resonance (EPR) tooth dosimetry for calculation of organ doses requires conversion of the measured absorbed dose in enamel. Before deriving conversion factors from simulation calculations with a realistic anthropomorphic human phantom, in the current study a simplified phantom was chosen to compare EPR measurement and Monte Carlo calculation. The dose response of tooth enamel of molars at various positions inside a cylindrical Plexiglas phantom of head-size was calculated by Monte Carlo modelling in parallel photon beams of X rays of 63 keV equivalent energy and 60Co gamma rays (1.25 Mev). For X ray exposure, preliminary results of EPR dosimetry with tooth enamel samples prepared from molars irradiated in the phantom were in agreement with calculation. The mean value of the ratio of the measured to the calculated dose was 0.93 ± 0.08.