Seemantini K. Nadkarni

Seemantini K. Nadkarni

Auto-generated profile, based on publicly available information

Seemantini K. Nadkarni

Wellman Center for Photomedicine, Massachusetts General Hospital, Harvard Medical School, Boston, USA

Research Focus & Metrics

Academic Career Length

14Years (2013-Present)

(Calculated once Scilit publication data is synced: current year − year of first publication + 1)

Representative Publications

View All Publications
24 February 2014 in Biomedical Optics Express
Verified
Journal Article
Open access
PDF file
MTMarkandey M. Tripathi
Markandey M. Tripathi
ZHZeinab Hajjarian
Zeinab Hajjarian
ECElizabeth M. Van Cott
Elizabeth M. Van Cott
SNSeemantini K. Nadkarni
Seemantini K. Nadkarni
We have developed and investigated a novel optical approach, Laser Speckle Rheology (LSR), to evaluate a patient’s coagulation status by measuring the viscoelastic properties of blood during coagulation. In LSR, a blood sample is illuminated with laser light and temporal speckle intensity fluctuations are measured using a high-speed CMOS camera. During blood coagulation, changes in the viscoelastic properties of the clot restrict Brownian displacements of light scattering centers within the sample, altering the rate of speckle intensity fluctuations. As a result, blood coagulation status can be measured by relating the time scale of speckle intensity fluctuations with clinically relevant coagulation metrics including clotting time and fibrinogen content. Our results report a close correlation between coagulation metrics measured using LSR and conventional coagulation results of activated partial thromboplastin time, prothrombin time and functional fibrinogen levels, creating the unique opportunity to evaluate a patient’s coagulation status in real-time at the point of care.

1 December 2016 in Scientific Reports
Verified
Journal Article
Open access
PDF file
ZHZeinab Hajjarian
Zeinab Hajjarian
HNHadi Tavakoli Nia
Hadi Tavakoli Nia
SAShawn Ahn
Shawn Ahn
AGAlan J. Grodzinsky
Alan J. Grodzinsky
RJRakesh K. Jain
Rakesh K. Jain
SNSeemantini K. Nadkarni
Seemantini K. Nadkarni
Natural and synthetic hydrogel scaffolds exhibit distinct viscoelastic properties at various length scales and deformation rates. Laser Speckle Rheology (LSR) offers a novel, non-contact optical approach for evaluating the frequency-dependent viscoelastic properties of hydrogels. In LSR, a coherent laser beam illuminates the specimen and a high-speed camera acquires the time-varying speckle images. Cross-correlation analysis of frames returns the speckle intensity autocorrelation function, g2(t), from which the frequency-dependent viscoelastic modulus, G*(ω), is deduced. Here, we establish the capability of LSR for evaluating the viscoelastic properties of hydrogels over a large range of moduli, using conventional mechanical rheometry and atomic force microscopy (AFM)-based indentation as reference-standards. Results demonstrate a strong correlation between |G*(ω)| values measured by LSR and mechanical rheometry (r = 0.95, p < 10−9), and z-test analysis reports that moduli values measured by the two methods are identical (p > 0.08) over a large range (47 Pa – 36 kPa). In addition, |G*(ω)| values measured by LSR correlate well with indentation moduli, E, reported by AFM (r = 0.92, p < 10−7). Further, spatially-resolved moduli measurements in micro-patterned substrates demonstrate that LSR combines the strengths of conventional rheology and micro-indentation in assessing hydrogel viscoelastic properties at multiple frequencies and small length-scales.

1 March 2020 in JACC: Cardiovascular Imaging
Verified
Journal Article
Closed access
PDPallavi Doradla
Pallavi Doradla
KOKenichiro Otsuka
Kenichiro Otsuka
ANAbhijay Nadkarni
Abhijay Nadkarni
MVMartin Villiger
Martin Villiger
AKAntonios Karanasos
Antonios Karanasos
SNSeemantini K. Nadkarni
Seemantini K. Nadkarni

1 May 2020 in Journal of Biomedical Optics
Verified
Journal Article
Open access
PDF file
ZHZeinab Hajjarian
Zeinab Hajjarian
SNSeemantini K. Nadkarni
Seemantini K. Nadkarni
The pathogenesis and progression of disease is frequently associated with serial changes in the mechanical properties of multiple functional units from organs and tissues to the extracellular matrix (ECM), cells, and subcellular structures.1 – 3 Biological tissues are viscoelastic, exhibiting both solid-like (elastic) and liquid-like (viscous) behaviors, under different loading conditions. This mechanical behavior is best quantified by the frequency-dependent shear viscoelastic modulus, G*(ω)=G′(ω)+iG′(ω).4 Here, the real part, G′(ω) is the elastic or storage modulus, a measure of solid-like behavior of the sample. On the other hand, the imaginary part, G″(ω) is the viscous or loss modulus and is a measure of viscous energy dissipation by the sample.4 Shear modulus varies markedly among biological tissues in different organs and is closely linked to the tissue function.1,5,6 This metric puts a number on the qualitative feeling of hardness, assessed by clinicians via physical palpation.5 Mechanical rheometry is the standard approach for evaluating the shear modulus, in which the specimen is subjected to an oscillatory shearing between two plates, and the ratio of applied stress to the resulting strain is measured over a limited oscillation frequency range.

23 August 2017 in Scientific Reports
Verified
Journal Article
Open access
PDF file
MTMarkandey M. Tripathi
Markandey M. Tripathi
SESatoru Egawa
Satoru Egawa
AWAlexandra G. Wirth
Alexandra G. Wirth
DTDiane M. Tshikudi
Diane M. Tshikudi
ECElizabeth M. Van Cott
Elizabeth M. Van Cott
SNSeemantini K. Nadkarni
Seemantini K. Nadkarni
Prothrombin time (PT) and the associated international normalized ratio (INR) are routinely tested to assess the risk of bleeding or thrombosis and to monitor response to anticoagulant therapy in patients. To measure PT/INR, conventional coagulation testing (CCT) is performed, which is time-consuming and requires the separation of cellular components from whole blood. Here, we report on a portable and battery-operated optical sensor that can rapidly quantify PT/INR within seconds by measuring alterations in the viscoelastic properties of a drop of whole blood following activation of coagulation with thromboplastin. In this study, PT/INR values were measured in 60 patients using the optical sensor and compared with the corresponding CCT values. Our results report a close correlation and high concordance between PT/INR measured using the two approaches. These findings confirm the accuracy of our optical sensing approach for rapid PT/INR testing in whole blood and highlight the potential for use at the point-of-care or for patient self-testing.

3 April 2014 in Journal of Biophotonics
Verified
Journal Article
Closed access
PDF file
ZHZeinab Hajjarian
Zeinab Hajjarian
MTMarkandey M. Tripathi
Markandey M. Tripathi
SNSeemantini K. Nadkarni
Seemantini K. Nadkarni