The Fourier theorem states that any time-series can be decomposed into a set of sinusoidal frequencies, each with its own phase and amplitude. The literature suggests that some frequencies are important to reproduce key qualities of eye-movements (“signal”) and some of frequencies are not important (“noise”). To investigate what is signal and what is noise, we analyzed our dataset in three ways: (1) visual inspection of plots of saccade, microsaccade and smooth pursuit exemplars; (2) analysis of the percentage of variance accounted for (PVAF) in 1,033 unfiltered saccade trajectories by each frequency band; (3) analyzing the main sequence relationship between saccade peak velocity and amplitude, based on a power law fit. Visual inspection suggested that frequencies up to 75 Hz are required to represent microsaccades. Our PVAF analysis indicated that signals in the 0-25 Hz band account for nearly 100% of the variance in saccade trajectories. Power law coefficients (a, b) return to unfiltered levels for signals low-pass filtered at 75 Hz or higher. We conclude that to maintain eye- movement signal and reduce noise, a cutoff frequency of 75 Hz is appropriate. We explain why, given this finding, a minimum sampling rate of 750 Hz is suggested.
In a prior report (Raju et al., 2023) we concluded that, if the goal was to preserve events such as saccades, microsaccades, and smooth pursuit in eye-tracking recordings, data with sine wave frequencies less than 75 Hz were the signal and data above 75 Hz were noise. Here, we compare five filters in their ability to preserve signal and remove noise. We compared the proprietary STD and EXTRA heuristic filters provided by our EyeLink 1000 (SR-Research, Ottawa, Canada), a Savitzky-Golay (SG) filter, an infinite impulse response (IIR) filter (low-pass Butterworth), and a finite impulse filter (FIR). For each of the non-heuristic filters, we systematically searched for optimal parameters. Both the IIR and the FIR filters were zero-phase filters. All filters were evaluated on 216 fixation segments (256 samples), from nine subjects. Mean frequency response profiles and amplitude spectra for all five filters are provided. Also, we examined the effect of our filters on a noisy recording. Our FIR filter had the sharpest roll-off of any filter. Therefore, it maintained the signal and removed noise more effectively than any other filter. On this basis, we recommend the use of our FIR filter. We also report on the effect of these filters on temporal autocorrelation.