In this study, a novel partially parallel acquisition (PPA) method is presented which can be used to accelerate image acquisition using an RF coil array for spatial encoding. This technique, GeneRalized Autocalibrating Partially Parallel Acquisitions (GRAPPA) is an extension of both the PILS and VD‐AUTO‐SMASH reconstruction techniques. As in those previous methods, a detailed, highly accurate RF field map is not needed prior to reconstruction in GRAPPA. This information is obtained from several k‐space lines which are acquired in addition to the normal image acquisition. As in PILS, the GRAPPA reconstruction algorithm provides unaliased images from each component coil prior to image combination. This results in even higher SNR and better image quality since the steps of image reconstruction and image combination are performed in separate steps. After introducing the GRAPPA technique, primary focus is given to issues related to the practical implementation of GRAPPA, including the reconstruction algorithm as well as analysis of SNR in the resulting images. Finally, in vivo GRAPPA images are shown which demonstrate the utility of the technique. Magn Reson Med 47:1202–1210, 2002.
In this study a novel partially parallel acquisition method is presented, which can be used to accelerate image acquisition using an RF coil array for spatial encoding. In this technique, Parallel Imaging with Localized Sensitivities (PILS), it is assumed that the individual coils in the array have localized sensitivity patterns, in that their sensitivity is restricted to a finite region of space. Within the PILS model, a detailed, highly accurate RF field map is not needed prior to reconstruction. In PILS, each coil in the array is fully characterized by only two parameters: the center of coil's sensitive region in the FOV and the width of the sensitive region around this center. In this study, it is demonstrated that the incorporation of these coil parameters into a localized Fourier transform allows reconstruction of full FOV images in each of the component coils from data sets acquired with a reduced number of phase encoding steps compared to conventional imaging techniques. After the introduction of the PILS technique, primary focus is given to issues related to the practical implementation of PILS, including coil parameter determination and the SNR and artifact power in the resulting images. Finally, in vivo PILS images are shown which demonstrate the utility of the technique. Magn Reson Med 44:602–609, 2000.