Non-invasive vital signs monitoring, especially blood pressure (BP), is crucial for evaluating overall health and identifying early indicators of medical issues. Photoplethysmography (PPG) is a non-invasive technique increasingly adopted in vital sign monitoring, as it allows for the continuous and precise measurement of cardiac-related signals. This technology is sensitive to fluctuations in blood volume within blood vessels, enabling real-time tracking of each heartbeat and identification of any irregularities. This study investigates the effectiveness of machine learning models for predicting arterial pressure based solely on PPG signals, utilizing a novel dataset and multi-wavelength sensor technology that captures data from four distinct wavelengths: infrared, red, green, and blue. The participant pool consisted of 88 individuals, with each measurement lasting 30 seconds at a sampling rate of 100 Hz. After a dedicated phase of signal preprocessing, three methodologies were assessed: a multi-layer perceptron (MLP) utilizing 84 features per subject, a dimensionality reduction strategy through Principal Component Analysis (PCA), and a Convolutional Neural Network (CNN) architecture. The CNN approach showcased impressive performance metrics, with results aligning with the standards set by the Association for the Advancement of Medical Instrumentation (AAMI) and the British Hypertension Society (BHS). Additionally, to further enhance accuracy, a Support Vector Regression (SVR) correction algorithm was applied, as a post-processing phase, addressing discrepancies in measurements related to age, thereby improving the reliability of the arterial pressure predictions and reaching errors of 1.36 ± 2.98 mmHg and 1.52 ± 2.67 mmHg for systolic and diastolic blood pressure respectively. Compared with recent cuff-less BP estimators that rely on single-wavelength PPG or PTT features and typically report MAE values of 3–5 mmHg, the proposed multi-wavelength hybrid CNN-SVR shrinks the error by more than 60% (MAE = 1.84/1.98 mmHg for SBP/DBP, Table II), while still meeting AAMI and BHS grade A/B requirements. This establishes, to the best of our knowledge, the first sub-2 mmHg, AAMI-compliant result obtained with a commodity optical front-end and no calibration.