1 July 2008 in Analytica Chimica Acta
1 January 2007 in Optics Express
We theoretically and experimentally analyze the biomolecule detection capability of the liquid core optical ring resonator (LCORR) as a label-free bio/chemical sensor. We first establish a simple and general linear relationship between the LCORR’s bulk refractive index sensitivity (BRIS) and its response to molecule deposition onto the surface, which enables us to easily characterize the LCORR sensing performance. Then, biosensing experiments are performed with bovine serum albumin (BSA) and LCORRs of various BRISs. The experimental results are in good agreement with the theoretical prediction. Further analysis shows that the LCORR is capable of detecting BSA below 10 pM with sub-picogram/mm2 mass detection limit.
1 February 2008 in Biosensors and Bioelectronics
8 January 2008 in The Analyst
We have demonstrated sensitive label-free virus detection using the opto-fluidic ring resonator (OFRR) sensor. The OFRR is a novel sensing platform that integrates the microfluidics and photonic sensing technology with a low detection limit and small volume. In our experiment, filamentous bacteriophage M13 was used as a safe model system. Virus samples were flowed through the OFRR whose surface was coated with M13-specific antibodies. We studied the sensor performance by monitoring in real-time the virus and antibody interaction. It is shown that OFRR can detect M13 with high specificity and sensitivity. The detection limit is approximately 2.3 × 103 pfu mL−1 and the detection dynamic range spanned seven orders of magnitude. Theoretical analysis was also carried out to confirm the experimental results. Our study will lead to development of novel OFRR-based, sensitive, rapid, and low-cost micro total analysis devices for virus detection.
26 December 2006 in IEEE Sensors Journal
NHNiranjan M. Hanumegowda
+1
We demonstrate refractive index measurement of liquids using two sensor system designs, both based on microring resonators. Evanescent sensors based on microrings utilize the resonating nature of the light to dramatically decrease the required size and sample consumption volume, which are requirements of lab-on-a-chip sensor systems. The first design, which utilizes an optical microsphere, exhibits a sensitivity of 30 nm/RIU and a resulting detection limit on the order of 10-7 RIU. The second approach is a novel design called a liquid core optical ring resonator (LCORR). This concept uses a quartz capillary as the fluidics and as the ring resonator and achieves a sensitivity of 16.1 nm/RIU. The detection limit of this system is around 5times10-6 RIU. Both of these systems have the potential to be incorporated with advanced microfluidic systems for lab-on-a-chip applications. In particular, the LCORR combines high sensitivity, performance stability, and microfluidic compatibility, making it an excellent choice for lab-on-a-chip development
19 December 2006 in Analytical Chemistry