This paper presents an apparatus for the calibration of long-stem platinum resistance thermometers at the argon triple point \((T_{90} = 83.8058\,\hbox {K})\) , designed at the Institute of Low Temperature and Structural Research, Poland (INTiBS). A hermetically sealed cell filled at the Istituto Nazionale di Ricerca Metrologica, Italy with high purity gas (6N) is the main element of this apparatus. The cell is placed in a cryostat fully immersed in liquid nitrogen. A temperature-controlled shield ensures the quasi-adiabatic condition needed for proper realization of the phase transition. A system for correcting the temperature distribution along the thermometer well is also implemented. The cell cooling and argon solidification is carried out by filling the thermometer well with liquid nitrogen. A LabVIEW computer program written at INTiBS automatically controls the triple-point realization process. The duration of a melting plateau in the apparatus lasts for about 24 h. The melting width for \(F\) between 20 % and 80 % was \({<}0.3\) mK. The reproducibility of the plateau temperature is better than \({\pm }20\,\upmu \hbox {K}\) .
13 April 2017 in International Journal of Thermophysics
The argon triple point (\(T_{90} = 83.8058\,\hbox {K}\)) is a fixed point of the International Temperature Scale of Preston-Thomas (Metrologia 27:3, 1990 ). Cells for realization of the fixed point have been manufactured by several European metrology institutes (Pavese in Metrologia 14:93, 1978 ; Pavese et al. in Temperature, part 1, American Institute of Physics, College Park, 2003 ; Hermier et al. in Temperature, part 1, American Institute of Physics, College Park, 2003 ; Pavese and Beciet in Modern gas-based temperature and pressure measurement, Springer, New York, 2013 ). The Institute of Low Temperature and Structure Research has in its disposal a few argon cells of various constructions used for calibration of capsule-type standard platinum resistance thermometers (CSPRT) that were produced within 40 years. These cells differ in terms of mechanical design and thermal properties, as well as source of gas filling the cell. This paper presents data on differences between temperature values obtained during the realization of the triple point of argon in these cells. For determination of the temperature, a heat-pulse method was applied (Pavese and Beciet in Modern gas-based temperature and pressure measurement, Springer, New York, 2013 ). The comparisons were performed using three CSPRTs. The temperatures difference was determined in relation to a reference function \(W(T)=R(T_{90})/R(271.16\hbox {K})\) in order to avoid an impact of CSPRT resistance drift between measurements in the argon cells. Melting curves and uncertainty budgets of the measurements are given in the paper. A construction of measuring apparatus is also presented in this paper.