Objectives: This study aimed to verify the enzymatic ethanol analysis method and compare the technique with headspace gas chromatography (HS/GC). Methods: The limit of the blank, limit of detection, and limit of quantification, precision, linearity, and accuracy of the enzymatic process were evaluated based on the data obtained. In addition, a method comparison was made between the enzymatic method and the HS/GC analysis method. Results: It was seen that the enzymatic method was linear in the range of 14–450 mg/dL, the coefficient of variation and bias values ranged between 0.5–1.09% and 4.52–9.78%, respectively. The quantitative accuracy was 2.13% and 2.23% at concentrations of 41.16 and 95.82 mg/dL, respectively. When the regression analysis was performed between the results obtained using the enzymatic alcohol dehydrogenase (ADH) and HS/GC methods, y=0.966x + 2.084 linear relationship formula and regression coefficient of r 2 =0.9987, it was found that it showed a good agreement at a confidence interval of 5%. Conclusions: The findings showed that the enzymatic method had good linearity, excellent precision, and a high correlation with the higher-order comparator method, the HS/GC method, in the ethanol analysis.