Vacuum pump wear is the most prevalent failure mode of the IBA Synthera® automated radiochemistry system. Rebuilding or replacing the pump causes equipment downtime and increases the radiation exposure of the service personnel. We built a dedicated test device to assess new or rebuilt pumps prior to installation, thus reducing downtime and radiation exposure during repairs. The Testbed incorporates a microprocessor that actuates the pump, valves, pressure sensor, and communicates with the user through lights, buttons, and an alphanumeric screen. The Testbed increases productivity and safety in the radiochemistry laboratory.

6 April 2023 in Instruments
Verified
Journal Article
Open access
PDF file
Victor Amador DiazVictor Amador Diaz
Victor Amador Diaz
SSScott E. Snyder
Scott E. Snyder
AVAmy L. Vavere
Amy L. Vavere
Vacuum pump wear is the most prevalent failure mode of the IBA Synthera® automated radiochemistry system. Rebuilding or replacing the pump causes equipment downtime and increases the radiation exposure of the service personnel. We built a dedicated test device to assess new or rebuilt pumps prior to installation, thus reducing downtime and radiation exposure during repairs. The Testbed incorporates a microprocessor that actuates the pump, valves, and pressure sensor; communicates with the user through lights, buttons, and an alphanumeric screen; and outputs test results to a laptop. The Testbed increases productivity and safety in the radiochemistry laboratory.