19 July 2026 in SSRN Electronic Journal
Levated urinary levels of interleukin-1 α (IL-1α) and vascular endothelial growth factor (VEGF) have been reported in patients with interstitial cystitis (IC) relative to other inflammatory bladder diseases, highlighting their potential as biomarkers for disease diagnosis and therapeutic monitoring. No compact platform for the concurrent detection of these two IC-associated biomarkers in a urine matrix has been reported to date. Here, we present a single-chip multiplexed electrochemical biosensor for the simultaneous quantification of IL-1α and VEGF in artificial urine. The device was fabricated on a glass substrate using a combination of e-beam evaporation and screen-printing techniques. It exhibited dynamic detection ranges of 1–1,000 pg/mL for VEGF and 0.01–100 pg/mL for IL-1α, with limits of detection of 10 fg/mL and 1 fg/mL and response times of 20 s and 15 s, respectively. The biosensor demonstrated reproducibility with a relative standard deviation (RSD) below 3%, selectivity against common urinary interferents, including 0.5 mM glucose and 80 nM nitrite, and no measurable cross-reactivity between sensing channels. Repeatability testing across four consecutive calibration runs yielded an RSD below 0.1%. The platform further exhibited reversible responses over four ascending–descending concentration cycles, minimal signal drift during 12 h of continuous operation (overall RSD ≈ 0.6%), and stable performance over seven days (overall RSD = 1%). The capability to quantitatively monitor IL-1α and VEGF in real time using a single analytical platform supports its potential application in IC diagnosis, disease progression monitoring, and evaluation of therapeutic response at the point of care.