We developed a monolateral external fixation system for bone elongation easier to assemble than commercial systems since the clamps can be laterally assembled to the rail in contrast to the commercial systems where the clamps can only be assembled from the ends. The system is formed by clamps, screws, and a rail. The rail is composed of two carbon fiber circular rods of ½” diameter and 42-cm length. The clamps can slide with respect to the rail to allow bone distraction or may be fixed to the rail by using a wedge and a screw. The prototype was mechanically evaluated under the ASTM F1541 standard. The stiffness in axial, Medio Lateral bending (ML), Antero Posterior bending (AP), and flexo-torsional load were 117 N/mm, 46 N/mm, 8.7 N/mm, and 5.8 N-mm/°, respectively, which are on the range of stiffness of commercial systems.
Currently, the cardiac dynamic of the entire cycle is not characterized and heart function is approximated by volumetric relationships in specific moments of the cycle. This work represents the left ventricular cardiac dynamics by intraventricular motion patterns in echocardiographic images. Briefly, after segmenting the ventricular cavity, intensity levels within this region of interest are spatially grouped by a classic superpixel (SP) clustering method, aiming to maximize the intergroup variance with a minimum number of groups. The distribution of these superpixels is then correlated with different cycle times, specifically diastole and systole. A total of 70 cases, half with Ejection Fraction (EF) less than 40 % and half with control EF (higher than 50 %), were processed and results demonstrated the number of groups is significantly smaller in cases with control EF, indicating that intraventricular patterns are correlated with the EF.