26 October 2015 in IEEE Transactions on Biomedical Circuits and Systems
Wearable devices offer interesting features, such as low cost and user friendliness, but their use for medical applications is an open research topic, given the limited hardware resources they provide. In this paper, we present an embedded solution for real-time EMG-based hand gesture recognition. The work focuses on the multi-level design of the system, integrating the hardware and software components to develop a wearable device capable of acquiring and processing EMG signals for real-time gesture recognition. The system combines the accuracy of a custom analog front end with the flexibility of a low power and high performance microcontroller for on-board processing. Our system achieves the same accuracy of high-end and more expensive active EMG sensors used in applications with strict requirements on signal quality. At the same time, due to its flexible configuration, it can be compared to the few wearable platforms designed for EMG gesture recognition available on market. We demonstrate that we reach similar or better performance while embedding the gesture recognition on board, with the benefit of cost reduction. To validate this approach, we collected a dataset of 7 gestures from 4 users, which were used to evaluate the impact of the number of EMG channels, the number of recognized gestures and the data rate on the recognition accuracy and on the computational demand of the classifier. As a result, we implemented a SVM recognition algorithm capable of real-time performance on the proposed wearable platform, achieving a classification rate of 90%, which is aligned with the state-of-the-art off-line results and a 29.7 mW power consumption, guaranteeing 44 hours of continuous operation with a 400 mAh battery.
4 April 2018 in IEEE Journal of Solid-State Circuits
We report VivoSoC, a system-on-chip realized in 130-nm CMOS for miniaturized medical instrumentation as used in mobile health devices or implantable telemetry systems for animal experiments. It features six neural stimulation channels and acquisition circuits for 9× electrode-based recordings, 4-channel/32-LED photoplethysmography (PPG), bioimpedance, and temperature. A 34-μW/MHz quad-core processing unit with sophisticated data, power, and clock management techniques enables on-chip feature extraction, which can dramatically reduce transmission or storage data rates in a system-crucial for overall power consumption and exemplarily demonstrated on preliminary work on implantable PPG-based vital signs monitoring.
1 November 2015 in IEEE Transactions on Circuits and Systems I: Regular Papers
In battery-powered medical instrumentation, the resolution and signal bandwidth of analog-to-digital converters (ADCs) have to be adapted to the needs of the application to avoid power wastage. This paper presents a reconfigurable successive approximation register (SAR) ADC implemented in 130 nm CMOS that resolves 5-14 bit with a maximum achievable effective number of bits (ENOB) of 13.5 using non-subtractive dither. In the proposed ADC design, the power consumption can be traded for accuracy to improve the energy efficiency and extend its application range, while reducing system integration complexity. A figure-of-merit (FoM) of 59 fJ/conversion is achieved at 1.2 V supply and the converter occupies an area of 0.42 mm 2 . Measurement results of the ADC integrated in a multi-channel analog front-end (AFE) circuit show the suitability of the ADC for portable medical monitoring devices.
12 January 2021 in IEEE Journal of Solid-State Circuits
This article presents a power-efficient low-jitter fractional-N digital phase-locked loop (DPLL) that resolves phase error (PE) in the fully differential voltage (FDV) domain. Compared with adopting a traditional time-to-digital converter (TDC), which relies on gate delay in the time domain, power-efficient quantization of PE by the proposed conversion scheme in the FDV domain can be accomplished with a higher power-supply/common-mode rejection ratio (PSRR/CMRR), lower process, voltage, and temperature (PVT) sensitivity, finer resolution, and better linearity. The implemented DPLL covers the fractional-N operation by a 10-bit differential current digital-to-analog converter (DAC) with a resistive load to represent the fractional phase/time. A differential dv/dt ramp is employed to linearly transfer the preset initial voltage into a small phase/voltage error, which is digitized by a narrow-range fine-resolution 7-bit self-timed successive-approximation-register analog-to-digital converter (SAR-ADC). The prototype DPLL, implemented in 130-nm CMOS, achieves 101-fs rms jitter, integrated from 10 kHz to 40 MHz, in the fractional-N mode with sub-12-bit fractional-frequency-control words, using an 80-MHz reference clock (REF), consuming 9.2 mW. This corresponds to a figure of merit (FoM) of -250.3 dB. The measured worst case fractional spur level is -56 dBc.