To date, the limited degrees of freedom (DOF) of most robotic training devices hinders them from providing functional training following stroke. We developed a 6-DOF exoskeleton (“BONES”) that allows movement of the upper limb to assist in rehabilitation. The objectives of this pilot study were to evaluate the impact of training with BONES on function of the affected upper limb, and to assess whether multijoint functional robotic training would translate into greater gains in arm function than single joint robotic training also conducted with BONES.
This paper describes a robotic-arm exoskeleton that uses a parallel mechanism inspired by the human forearm to allow naturalistic shoulder movements. The mechanism can produce large forces through a substantial portion of the range of motion (RoM) of the human arm while remaining lightweight. This paper describes the optimization of the exoskeleton's torque capabilities by the modification of the key geometric design parameters.
23 November 2016 in IEEE/ASME Transactions on Mechatronics
This paper describes how parallel elastic elements can be used to reduce energy consumption in the electric-motor-driven, fully actuated, Sandia Transmission-Efficient Prototype Promoting Research (STEPPR) bipedal walking robot without compromising or significantly limiting locomotive behaviors. A physically motivated approach is used to illustrate how selectively engaging springs for hip adduction and ankle flexion predict benefits for three different flat-ground walking gaits: human walking, human-like robot walking, and crouched robot walking. Based on locomotion data, springs are designed and substantial reductions in power consumption are demonstrated using a bench dynamometer. These lessons are then applied to STEPPR, a fully actuated bipedal robot designed to explore the impact of tailored joint mechanisms on walking efficiency. Featuring high-torque brushless DC motors, efficient low-ratio transmissions, and high-fidelity torque control, STEPPR provides the ability to incorporate novel joint-level mechanisms without dramatically altering high-level control. Unique parallel elastic designs are incorporated into STEPPR, and walking data show that hip adduction and ankle flexion springs significantly reduce the required actuator energy at those joints for several gaits. These results suggest that parallel joint springs offer a promising means of supporting quasi-static joint torques due to body mass during walking, relieving motors of the need to support these torques and substantially improving locomotive energy efficiency.
18 April 2012 in IEEE Transactions on Neural Systems and Rehabilitation Engineering
Training with haptic guidance has been proposed as a technique for learning complex movements in rehabilitation and sports, but it is unclear how to best deliver guidance-based training. Here, we hypothesized that breaking down a complex movement, similar to a tennis backhand, into simpler parts and then using haptic feedback from a robotic exoskeleton would help the motor system learn the movement. We also examined how the particular form of the decomposition affected learning. Three groups of unimpaired participants trained with the target arm movement broken down in three ways: 1) elbow flexion/extension and the unified shoulder motion independently (“anatomical” decomposition), 2) three component shoulder motions in Euler coordinates and elbow flexion/extension (“Euler” decomposition), or 3) the motion of the tip of the elbow and motion of the hand with respect to the elbow, independently (“visual” decomposition). A control group practiced the same number of movements, but experienced the target motion only, achieving eight times more direct practice with this motion. Despite less experience with the target motion, part training was better, but only when the arm trajectory was decomposed into anatomical components. Varying robotic movement training to include practice of simpler, anatomically-isolated motions may enhance its efficacy.
This paper presents a novel design for a 4 degree of freedom pneumatically-actuated upper-limb rehabilitation device. BONES is based on a parallel mechanism that actuates the upper arm by means of two passive, sliding rods pivoting with respect to a fixed structural frame. Four, mechanically-grounded pneumatic actuators are placed behind the main structural frame to control shoulder motion via the sliding rods, and a fifth cylinder is located on the structure to control elbow flexion/extension. The device accommodates a wide range of motion of the human arm, while also achieving low inertia and direct-drive force generation capability at the shoulder. A key accomplishment of this design is the ability to generate arm internal/external rotation without any circular bearing element such as a ring, a design feature inspired by the biomechanics of the human forearm. The paper describes the rationale for this device and its main design aspects including its kinematics, range of motion, and force generation capability.
The robot described in this paper, SUE (Supinator Extender), adds forearm/wrist rehabilitation functionality to the UCI BONES exoskeleton robot and to the ArmeoSpring rehabilitation device. SUE is a 2-DOF serial chain that can measure and assist forearm supination-pronation and wrist flexion-extension. The large power to weight ratio of pneumatic actuators allows SUE to achieve the forces needed for rehabilitation therapy while remaining lightweight enough to be carried by BONES and ArmeoSpring. Each degree of freedom has a range of 90 degrees, and a nominal torque of 2 ft-lbs. The cylinders are mounted away from the patient's body on the lateral aspect of the arm. This is to prevent the danger of a collision and maximize the workspace of the arm robot. The rotation axis used for supination-pronation is a small bearing just below the subject's wrist. The flexion-extension motion is actuated by a cantilevered pneumatic cylinder, which allows the palm of the hand to remain open. Data are presented that demonstrate the ability of SUE to measure and cancel forearm/wrist passive tone, thereby extending the active range of motion for people with stroke.