Adaptive deep brain stimulation (aDBS) is a promising concept for feedback-based neurostimulation, with the potential of clinical implementation with the sensing-enabled Percept neurostimulator. We aim to characterize chronic electrophysiological activity during stimulation and to validate beta-band activity as a biomarker for bradykinesia. Subthalamic activity was recorded during stepwise stimulation amplitude increase OFF medication in 10 Parkinson’s patients during rest and finger tapping. Offline analysis of wavelet-transformed beta-band activity and assessment of inter-variable relationships in linear mixed effects models were implemented. There was a stepwise suppression of low-beta activity with increasing stimulation intensity (p = 0.002). Low-beta power was negatively correlated with movement speed and predictive for velocity improvements (p < 0.001), stimulation amplitude for beta suppression (p < 0.001). Here, we characterize beta-band modulation as a chronic biomarker for motor performance. Our investigations support the use of electrophysiology in therapy optimization, providing evidence for the use of biomarker analysis for clinical aDBS.
Subthalamic beta band activity (13–35 Hz) is known as a real-time correlate of motor symptom severity in Parkinson’s disease (PD) and is currently explored as a feedback signal for closed-loop deep brain stimulation (DBS). Here, we investigate the interaction of movement, dopaminergic medication, and deep brain stimulation on subthalamic beta activity in PD patients implanted with sensing-enabled, implantable pulse generators. We recorded subthalamic activity from seven PD patients at rest and during repetitive movements in four conditions: after withdrawal of dopaminergic medication and DBS, with medication only, with DBS only, and with simultaneous medication and DBS. Medication and DBS showed additive effects in improving motor performance. Distinct effects of each therapy were seen in subthalamic recordings, with medication primarily suppressing low beta activity (13–20 Hz) and DBS being associated with a broad decrease in beta band activity (13–35 Hz). Movement suppressed beta band activity compared to rest. This suppression was most prominent when combining medication with DBS and correlated with motor improvement within patients. We conclude that DBS and medication have distinct effects on subthalamic beta activity during both rest and movement, which might explain their additive clinical effects as well as their difference in side-effect profiles. Importantly, subthalamic beta activity significantly correlated with motor symptoms across all conditions, highlighting its validity as a feedback signal for closed-loop DBS.
Beta-band activity in the subthalamic local field potential (LFP) is correlated with Parkinson’s disease (PD) symptom severity and is the therapeutic target of deep brain stimulation (DBS). While beta fluctuations in PD patients are well characterized on shorter timescales, it is not known how beta activity evolves around the diurnal cycle, outside a clinical setting. Here, we obtained chronic recordings (34 ± 13 days) of subthalamic beta power in PD patients implanted with the Percept DBS device during high-frequency DBS and analysed their diurnal properties as well as sensitivity to artifacts. Time of day explained 41 ± 9% of the variance in beta power (p < 0.001 in all patients), with increased beta during the day and reduced beta at night. Certain movements affected LFP quality, which may have contributed to diurnal patterns in some patients. Future DBS algorithms may benefit from taking such diurnal and artifactual fluctuations in beta power into account.
Brain sensing devices are approved today for Parkinson's, essential tremor, and epilepsy therapies. Clinical decisions for implants are often influenced by the premise that patients will benefit from using sensing technology. However, artifacts, such as ECG contamination, can render such treatments unreliable. Therefore, clinicians need to understand how surgical decisions may affect artifact probability. Investigate neural signal contamination with ECG activity in sensing enabled neurostimulation systems, and in particular clinical choices such as implant location that impact signal fidelity. Electric field modeling and empirical signals from 85 patients were used to investigate the relationship between implant location and ECG contamination. The impact on neural recordings depends on the difference between ECG signal and noise floor of the electrophysiological recording. Empirically, we demonstrate that severe ECG contamination was more than 3.2x higher in left-sided subclavicular implants (48.3%), when compared to right-sided implants (15.3%). Cranial implants did not show ECG contamination. Given the relative frequency of corrupted neural signals, we conclude that implant location will impact the ability of brain sensing devices to be used for “closed-loop” algorithms. Clinical adjustments such as implant location can significantly affect signal integrity and need consideration.
Background and purpose: Biomarkers for future adaptive deep brain stimulation still need evaluation in clinical routine. Here, we aimed to assess stimulation‐induced modulation of beta‐band activity and clinical symptoms in a Parkinson's disease patient during chronic neuronal sensing using a novel implantable pulse generator.Methods: Subthalamic activity was recorded OFF and ON medication during a stepwise increase of stimulation amplitude. Off‐line fast fourier transfom ‐based analysis of beta‐band activity was correlated with motor performance rated from blinded videos.Results: The stepwise increase of stimulation amplitude resulted in decreased beta oscillatory activity and improvement of bradykinesia. Mean low beta‐band (13–20 Hz) activity correlated significantly with bradykinesia (ρ = 0.662, p < 0.01).Conclusions: Motor improvement is reflected in reduced subthalamic beta‐band activity in Parkinson's disease, supporting beta activity as a reliable biomarker. The novel PERCEPT neurostimulator enables chronic neuronal sensing in clinical routine. Our findings pave the way for a personalized precision‐medicine approach to neurostimulation.
Deep brain stimulation (DBS) is an invasive treatment option for patients with Parkinson's disease. Recently, adaptive DBS (aDBS) systems have been developed, which adjust stimulation timing and amplitude in real-time. However, it is unknown how changes in parameters, movement states and the controllability of subthalamic beta activity affect aDBS performance. To characterize how parameter choice, movement state and controllability interactively affect the electrophysiological and behavioral response to single threshold aDBS. We recorded subthalamic local field potentials in 12 patients with Parkinson's disease receiving single threshold aDBS in the acute post-operative state. We investigated changes in two aDBS parameters: the onset time and the smoothing of real-time beta power. Electrophysiological patterns and motor performance were assessed while patients were at rest and during a simple motor task. We further studied the impact of controllability on aDBS performance by comparing patients with and without beta power modulation during continuous stimulation. Our findings reveal that changes in the onset time control the extent of beta power suppression achievable with single threshold adaptive stimulation during rest. Behavioral data indicate that only specific parameter combinations yield a beneficial effect of single threshold aDBS. During movement, action induced beta power suppression reduces the responsivity of the closed loop algorithm. We further demonstrate that controllability of beta power is a prerequisite for effective parameter dependent modulation of subthalamic beta activity. Our results highlight the interaction between single threshold aDBS parameter selection, movement state and controllability in driving subthalamic beta activity and motor performance. By this means, we identify directions for the further development of closed-loop DBS algorithms.