A method for the determination of lead in mussel slurries by atomic absorption spectrometry with electrothermal atomization using Pd–Mg(NO3)2 as a chemical modifier is proposed. The drying, ashing and atomization temperatures were optimized. Calibration and standard additions graphs, sensitivity, precision, accuracy and the effects of different slurry stabilization agents were also studied. The method was applied to the determination of lead in mussel samples from the Galician coast and the results obtained were compared with those achieved by means of a wet digestion procedure.
1 January 1993 in Analytical and Bioanalytical Chemistry
A knotted reactor flow injection on-line preconcentration system coupled to a flame atomic absorption spectrometer (FAAS) was developed for the direct determination of cadmium in aqueous samples. Diethyldithiocarbamate was used to complex the analyte and then, it was adsorbed on the walls of a knotted reactor (KR) made from PTFE tubing. All the FI and chemical variables were optimized. The complex elution was performed on-line using ethanol. The sample loading time and flow rate were set at 60 s and 6 mL/min, respectively. An enhancement factor of 33 was achieved, resulting in a sampling frequency (h−1) of 30. The limit of detection was 3.44 μg/L. The accuracy of the method was tested by the analysis of three certified reference materials (Certified Aqueous Standard 7878, Reference Water TM-24 and Trace Elements in Water 1643d). The results obtained are within the certified ranges at a 95% confidence level and a good precision was also achieved (< 2.7%, n = 6). The proposed procedure was applied to the Cd...