Circulating endothelial cells (CECs) are a reliable biomarker for cardiovascular diseases (CVDs). A major unresolved challenge limiting the widespread use of CECs for the diagnosis and monitoring of CVDs is their unreliable detection. This problem is mainly attributed to the low sample volume (5-10 ml) of commonly used ex vivo CEC isolation methods. To overcome this limitation, the BMProbeTM for the in vivo isolation of CECs is proposed. It consists of a twisted medical flat wire with a polymer-coated surface functionalized with anti-CD105 antibodies. A combined experimental and numerical study was performed to investigate which flow conditions lead to an increased cell attachment to the probe‘s surface. Endothelial cells were solved in a dextran solution and circulated in a flow system containing the BMProbesTM. Microscopic images of the attached CECs were taken. In addition, the in vitro experiments were simulated using a computational fluid dynamics (CFD) flow solver to quantify the flow conditions at the probe‘s surface. The microscopic images were superimposed with the CFD data to investigate the influence of wall shear rate and wall normal rate on the attachment of CECs to the probe. Most of all attached cells (85.5%) were found in areas of negative wall normal rate.