An 80-year-old woman presented to our hospital with presyncope. Holter monitoring revealed paroxysmal atrial fibrillation and asystole due to sick sinus syndrome. Twelve years ago, she underwent surgery for right breast cancer and received postoperative radiotherapy; the right atrium was included in the irradiation area. She was not taking antiarrhythmic drugs. At admission, her blood pressure was 133/72 mmHg. A 12-lead electrocardiogram revealed a heart rate of 76 beats/min and inverted P waves in leads II, III, and aVF, suggesting an ectopic atrial rhythm. Transthoracic echocardiography revealed normal left ventricular systolic function. Moderate mitral and tricuspid regurgitation and bilateral atrial enlargement were observed. The electrolyte levels were normal. At the time of pacemaker implantation, P-wave sensing could not be confirmed with the tined lead; therefore, it was replaced with a screw lead and placed in the anterior location of the tricuspid annulus after trying several positions (
Figures 1 and
Supplementary material).
1 P-wave sensing was low (0.3 mV); however, the pacing parameters were acceptable. During and after pacemaker implantation, atrial pacing latency was prolonged to approximately 300 ms from atrial pacing to atrial contraction (
Figure 2). There was no abnormality in the pacing threshold and lead resistance or lead position. No significant clinical events were documented at follow-up 10 months later.