1 October 2018 in Frontiers in Neuroscience
This study reports the realization of an optical cochlear implant (oCI) with optimized thermomechanical properties for optogenetic experiments. The oCI probe comprises 144 miniaturized light-emitting diodes (μLEDs) distributed along a bendable, 1.5-cm-long, 350-μm-wide and 26-μm-thick probe shaft, individually controlled via a n × p matrix interconnection. In contrast to our earlier approach based on polyimide (PI) and epoxy resin with different thermal expansion coefficients, the μLEDs and interconnecting wires are now embedded into a triple-layer stack of a single, biocompatible, and highly transparent epoxy material. The new material combination results in a pronounced reduction of thermomechanical bending in comparison with the material pair of the earlier approach. We developed a spin-coating process enabling epoxy resin layers down to 5 μm at thickness variations of less than 7% across the entire carrier wafer. We observed that the cross-linking of epoxy resin layers strongly depends on the spin-coating parameters which were found to be correlated to a potential separation of epoxy resin components of different densities. Furthermore, various metallization layers and corresponding adhesion promoting layers were investigated. We identified the combination of silicon carbide with a titanium-based metallization to provide the highest peeling strength, achieving an adhesion to epoxy improved by a factor of two. In order to obtain a high process yield, we established a stress-free implant release using the electrochemical dissolution of a sacrificial aluminum layer. The direct comparison of oCI probe variants using a single epoxy material and the combination of PI and epoxy resin revealed that the epoxy-resin-only probe shows minimal thermomechanical probe bending with a negligible hysteresis. The thermal probe characterization demonstrated that the temperature increase is limited to 1 K at μLED DC currents of up to 10 mA depending on the stimulation duration and the medium surrounding the probe. The optical output power and peak wavelengths of the new oCI variant were extracted to be 0.82 mW and 462 nm when operating the μLEDs at 10 mA, 10 kHz, and a duty cycle of 10%. The optical power corresponds to a radiant emittance of 407 mW/mm2, sufficient for optogenetic experiments using channelrhodopsin-2.
29 June 2020 in EMBO Molecular Medicine
Electrical cochlear implants (eCIs) partially restore hearing and enable speech comprehension to more than half a million users, thereby re‐connecting deaf patients to the auditory scene surrounding them. Yet, eCIs suffer from limited spectral selectivity, resulting from current spread around each electrode contact and causing poor speech recognition in the presence of background noise. Optogenetic stimulation of the auditory nerve might overcome this limitation as light can be conveniently confined in space. Here, we combined virus‐mediated optogenetic manipulation of cochlear spiral ganglion neurons (SGNs) and microsystems engineering to establish acute multi‐channel optical cochlear implant (oCI) stimulation in adult Mongolian gerbils. oCIs based on 16 microscale thin‐film light‐emitting diodes (μLEDs) evoked tonotopic activation of the auditory pathway with high spectral selectivity and modest power requirements in hearing and deaf gerbils. These results prove the feasibility of μLED‐based oCIs for spectrally selective activation of the auditory nerve.
13 January 2020 in IEEE Transactions on Biomedical Engineering
This paper reports on the development, characterization and in vivo validation of compact optical neural probes. These novel intracerebral devices comprise micro light-emitting diodes (μLEDs) integrated along their slender probe shanks with up to 20 μLEDs per device. Blue light with a peak wavelength of 455nm is emitted from circular apertures 100μm in diameter. The μLEDs are structured on GaN-on-sapphire wafers and subsequently transferred onto silicon (Si) carrier wafers. The wafer-scale transfer process provides the opportunity to process the functional GaN layer stack from both sides and hence enables maximizing the efficiency of the μLEDs. Combined with standard MEMS fabrication processes for Si, linear μLED arrays with small inter-μLED distances are achieved on thin probe shanks with cross-sections measuring 150μm × 65 μm. Devices are interconnected using highly flexible polyimide cables in order to mechanically decouple them from the peripheral electronics during in vivo experiments. Assembled probes emit a peak optical radiant flux of 440μW (emittance $56mWmm^{\minus2}$ ) at 5mA driving current. Thermal characterization of test probes reveals a temperature increase of 1.5K measured using an integrated thermistor. Electrical functionality stress tests have been carried out to evaluate the device passivation against the physiological environment. It is estimated to endure at least 48 h during continuously pulsed μLED operation. A compact driving circuitry enables low-noise μLED operation in in vivo optogenetic experiments. The radiant flux necessary to elicit an acceptable neuronal response is determined between 1.36μW and 17.5 μW. Probe validation successfully demonstrates the layer-specific stimulation in the cortex in multiple in vivo trials.
1 January 2020 in Thin Solid Films
8 July 2019 in Journal of Micromechanics and Microengineering
This paper reports on the optimization of wafer-scale bonding of micro-light-emitting diodes (µLEDs) on four-inch sapphire wafers onto silicon (Si) substrates using an indium–gold (In–Au) reflow process. The bonding yield was optimized from initial values below 10% to 100% by applying a systematic variation of the two most relevant bonding parameters, namely the indium (In) bonding pad height t 2 and the relative bonding area Q . The fabrication process is designed to reliably transfer mLEDs from sapphire to Si substrates. The process is demonstrated by realizing optoelectronic probes carrying arrays of mLEDs dedicated for applications in neuroscientific research applying optogenetic methods. The 6 m thick mLEDs comprise of a gallium nitride layer epitaxially grown on sapphire wafers and lithographically patterned into mLEDs with lateral dimensions down to using reactive ion etching. The mLEDs are transferred onto Si substrates carrying interconnecting tracks and bonding pads made of gold (Au). Two technological challenges were successfully addressed. The first is to pattern thick In layers with lateral dimensions down to on top of the 6 m high mLED mesas. The second results from residual thermomechanical stress after wafer bonding due to the thermal expansion mismatch of sapphire and Si. It is demonstrated that the stress induced rupture of the In–Au bond interface between sapphire and Si is completely suppressed for t 2 , Q combinations with m and . These results enable fully functional two-dimensional arrays with mLEDs to be realized on 65 m thin Si substrates.
22 December 2021 in IET Optoelectronics
This paper reports on new schemes based on the concept of tri‐state switching for routing linear arrays of light‐emitting diodes (LEDs) on slim substrates. The schemes use a minimal number of wires in single‐metal planar technologies, where wires are not allowed to cross. They have in common that the number of LEDs, N L , addressable by N W wires is given by N L = 4 N W − 6. The designs are built on a family of hierarchically interconnected structures with (2
n
+1
− 2) LEDs and (2
n
−1
+ 1) wires, for positive integers n . A process termed linear expansion straightforwardly extends them to arbitrary values of N W and adds 4 LEDs with each additional wire. Expressions for series resistances and the average wire length normalized to the array length are derived. In hierarchical designs, the average normalized wire length asymptotically approaches 7/12 for large n . A matrix formulation graphically elucidates the new interconnection schemes.