1 April 2021 in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
Recently, super-harmonic ultrasound imaging technology has caused much attention due to its capability of distinguishing microvessles from the tissues surrounding them. However, the fabrication of a dual-frequency confocal transducer is still a challenge. In this work, 270μm PMN-PT single crystal 1-3 composite and 28μm PVDF thick film, acting as transmission layer and receiving layer, respectively, are integrated in a novel co-focusing structure. To realize delicate wave propagation control, microwave transmission line theory is introduced to design such structure. Two acoustic filter layers, 13μm copper layer and 39μm Epoxy 301 layer, are indispensable and should be added between two piezoelectric layers. Therefore, an acoustic issue can be overcome via an electrical method and the successful achievement of a dual-frequency (5MHz/30MHz) ultrasound transducer with confocal distance of 8mm can be realized. The super harmonic ultrasound imaging experiment is conducted using this kind of device. 3D image of 110μm diameter phantom tube injected with microbubbles can be obtained. These promising results demonstrate that this novel dual-frequency (5MHz/30MHz) confocal ultrasound transducer is potentially usable for microvascular medical imaging application in the future.