This paper presents a strategy for achieving metrological traceability using vector network analyzers (VNAs) at submillimeter-wave frequencies (300 to 3000 GHz). The strategy includes the use of traceable calibration techniques designed for operation at these frequencies. Slight, but significant, physical differences between the waveguide line standards, used during calibration, are accommodated by applying a weighting technique to combine results using different calibration lines. Measurement uncertainty is assessed by analyzing replicate measurement data, to take account of the different waveguide interface interactions that occur when the line standards are connected to the VNA. The strategy is illustrated using measurements made in the WM-250 (750 GHz to 1100 GHz) waveguide band.
This paper presents a detailed assessment of two rectangular metallic waveguide lines in order that they can be used as primary standards to provide metrological traceability for electrical scattering parameter measurements at submillimetre wavelengths. The assessment comprises a series of dimensional measurements to determine the overall quality of the lines in terms of the waveguide aperture size and alignment. This is followed by electrical measurements to confirm the electrical behaviour of the lines. Finally, the lines are employed as standards to calibrate a vector network analyser which is used to measure devices to verify the performance of the lines as calibration standards, in operando. The waveguide size is WM-380, which operates from 500 GHz to 750 GHz.
This paper describes a new facility that has been introduced recently to provide high precision traceable scattering coefficient measurements of waveguide devices in the frequency range 110 GHz to 170 GHz (i.e. in waveguide size WR-06). The facility comprises measurement instrumentation situated at the University of Leeds and associated primary reference standards provided by the National Physical Laboratory. The instrumentation consists of a Vector Network Analyzer (VNA) and the standards are precision sections of waveguide that are used to calibrate the VNA. Traceability to national standards and the International System of units (SI) is achieved via precision dimensional measurements of the waveguide sections. Typical measurements, with uncertainties, are given to illustrate the current state-of-the-art for traceable measurements of this type.
23 February 2024 in IEEE Transactions on Instrumentation and Measurement
Multi-port microwave devices operating at cryogenic temperatures are used in quantum computing to enable complex quantum operations. These components need to be precisely characterized at their operating temperature to ensure adequate overall system performance. In this work, an S-parameter measurement system to characterize the performance of coaxial connectorized microwave devices having up to four ports at temperatures of 3 K and above is presented. A four-port calibration setup is implemented at the 3 K temperature stage inside a dilution refrigerator. Measurement results for devices under test at 3 K are presented and the performance of the measurement system is evaluated.
30 November 2021 in Journal of Electromagnetic Engineering and Science
This paper evaluates the uncertainty of S-parameter measurements on multilayer printed circuit boards (PCBs) due to the uncertainties of the dimensions and dielectric properties of the line standard in the Thru-Reflect-Line (TRL) calibration. This evaluation is performed in two ways: one is based on repeated TRL calibrations with a randomly perturbed line standard, and the other is based on equations given by Stumper. The two methods require the uncertainties of the S-parameters of the TRL line standard, which are obtained from the uncertainties of the dimensions and dielectric properties using three-dimensional electromagnetic Monte Carlo simulation. The two methods agree well with each other. This study also shows how to apply impedance renormalization in Stumper’s equations. We design the TRL standards and the devices under test (DUTs) in PCB stripline and precisely measure the cross-sectional dimensions of the fabricated striplines. Uncertainty analysis based on the measured values enables us to investigate the impact of realistic deviations in the dimensions of the TRL line standard on the S-parameter measurement uncertainty of the DUTs. Finally, as an example, we evaluated the uncertainty in the measured S-parameters of a Beatty line on the fabricated PCB.