Potentiometric redox measurements were made in solutions of increasing biological complexity starting with buffer solutions containing either potassium ferri/ferrocyanide or ascorbic acid and finishing with plasma and blood. When the concentration of ferri/ferrocyanide was high (∼0.2 mM), both biofouled planar and nanoporous gold electrodes gave Nernstian slopes of 55–59 mV. However, at or below a critical concentration (≤ 0.1 mM), ∼20% loss in sensitivity was observed at planar gold in contrast to nanoporous gold where Nernstian behavior was retained. For ascorbic acid, a Nernst slope of −41 mV was observed at biofouled nanoporous gold electrodes. In contrast, biofouled planar gold electrodes failed to give any potentiometric redox response. At all concentrations studied, cyclic voltammetric measurements on biofouled electrodes revealed significant impairment of faradaic electroactivity at planar gold electrodes while no impairment was shown at nanoporous gold. These results indicate that nanoporous gold is an ideal electrode material to use when making both potentiometric and cyclic voltammetric measurements, particularly in complex solutions containing relatively low concentrations of redox molecules. As proof of concept, the redox potential of plasma and blood has been measured using nanoporous and planar gold electrodes and their values compared.
Potentiometric redox sensing is a relatively inexpensive and passive approach to evaluate the overall redox state of complex biological and environmental solutions. The ability to make such measurements in ultra-small volumes using high surface area, nanoporous electrodes is of particular importance as such electrodes can improve the rates of electron transfer and reduce the effects of biofouling on the electrochemical signal. This work focuses on the fabrication of miniaturized nanoporous gold (NPG) electrodes with a high surface area and a small footprint for the potentiometric redox sensing of three biologically relevant redox molecules (ascorbic acid, uric acid, and cysteine) in microliter volumes. The NPG electrodes were inexpensively made by attaching a nanoporous gold leaf prepared by dealloying 12K gold in nitric acid to a modified glass capillary (1.5 mm id) and establishing an electrode connection with copper tape. The surface area of the electrodes was ~1.5 cm2, providing a roughness factor of ~16 relative to the geometric area of 0.09 cm2. Scanning electron microscopy confirmed the nanoporous framework. A linear dependence between the open-circuit potential (OCP) and the logarithm of concentration (e.g., Nernstian-like behavior) was obtained for all three redox molecules in 100 μL buffered solutions. As a first step towards understanding a real system, the response associated with changing the concentration of one redox species in the presence of the other two was examined. These results show that at NPG, the redox potential of a solution containing biologically relevant concentrations of ascorbic acid, uric acid, and cysteine is strongly influenced by ascorbic acid. Such information is important for the measurement of redox potentials in complex biological solutions.
Potentiometric redox sensing in solutions containing multiple redox molecules was evaluated using in-house constructed nanoporous gold (NPG)-platinum (Pt) and unmodified NPG electrodes. The NPG-Pt electrode was fabricated by electrodepositing Pt into the nanoporous framework of a chemically dealloyed NPG electrode. By varying the concentration of the Pt salt and the electrodeposition time, different amounts of Pt were introduced. Characterization by SEM shows the pore morphology doesn’t change with the addition of Pt and XPS indicates the electrodes contain ∼2.5–24 wt% Pt. Open-circuit potential (OCP) measurements in buffer and solutions containing ascorbic acid, cysteine, and/or uric acid show that the OCP shifts positive with the addition of Pt. These results are explained by an increase in the rate of the oxygen reduction reaction with the addition of Pt. The overall shape of the potentiometric titration curves generated from solutions containing one or more bioreagents is also highly dependent on the amount of Pt in the nanoporous electrode. Furthermore, the generation of OCP vs Log [bioreagent] from the results of the potentiometric experiments shows an ∼2-fold increase in sensitivity can result with the addition of Pt. These results indicate the promise that these electrodes have in potentiometric redox sensing.