Due to the physical principles of the energy deposition of charged particles in matter, cancer treatment using proton therapy allows more conformal dose deposition in the tumour with respect to conventional photon radiotherapy, which leads to better protection of healthy tissue. At the same time, the maximum dose deposition at the end of the proton trajectory is more sensitive to uncertainties in the range of the protons. A significant source of these uncertainties are inter-fractional anatomical changes in the patient during treatment. Adaptive radiotherapy, where the treatment plan is adapted to the anatomy of the day, is an important tool to optimise dose deposition. Pixelated semiconductor detectors with energy resolution allow the kinetic energy of the protons to be measured in two dimensions. If the detector is placed downstream of the patient, the energy loss in the patient, expressed in terms of their water-equivalent thickness (WET), can be quantised (proton radiography). In this way, the patient's anatomy in the target area can be monitored over the course of treatment with a very small additional imaging dose, allowing the treatment plan to be optimised if the anatomy changes significantly. In this paper we present a proton radiography system using ATLAS IBL FE-I4 detector modules. The particle flux and the deposited energy are well beyond the intended application of the readout chip. We show first results for the measurement of the proton energy and the WET resolution and demonstrate the applicability of the detector for the acquisition of proton radiography images.
Accurate knowledge of a patient’s anatomy during every treatment fraction in proton therapy is an important prerequisite to ensure a correct dose deposition in the target volume. Adaptive proton therapy aims to detect those changes and adjust the treatment plan accordingly. One way to trigger a daily re-planning of the treatment is to take a proton radiograph from the beam’s-eye view before the treatment to check for possible changes in the water equivalent thickness (WET) along the path due to daily changes in the patient’s anatomy. In this paper, the Two-Plane Imaging System (TPIS) is presented, comprising two ATLAS IBL silicon pixel-detector modules developed for the tracking detector of the ATLAS experiment at CERN. The prototype of the TPIS is described in detail, and proof-of-principle WET images are presented, of two-step phantoms and more complex phantoms with bone-like inlays (WET 10 to 40 mm). This study shows the capability of the TPIS to measure WET images with high precision. In addition, the potential of the TPIS to accurately determine WET changes over time down to 1 mm between subsequently taken WET images of a changing phantom is shown. This demonstrates the possible application of the TPIS and ATLAS IBL pixel-detector module in adaptive proton therapy.