We have measured the resistance of a superconducting transition edge sensor over the entire transition region on a fine scale, producing a four-dimensional map of the resistance surface. The dimensionless temperature and current sensitivities ( and ) of the TES resistance have been determined at each point. α and β are closely related to the sensor performance, but show a great deal of complex, large amplitude fine structure over large portions of the surface that is sensitive to the applied magnetic field. We discuss the relation of this structure to the presence of Josephson “weak link” fringes.
23 April 2024 in Journal of Low Temperature Physics
Superconducting/normal metal bilayers with tunable transition temperature are a critical ingredient to the fabrication of high-performance transition edge sensors. Popular material choices include Mo/Au and Mo/Cu, which exhibit good environmental stability and provide low resistivity films to achieve adequate thermal conductivity. The deposition of high-quality Mo films requires sufficient adatom mobility, which can be provided by energetic ions in sputter deposition or by heating the substrate in an e-beam evaporation process. The bilayer Tc depends sensitively on the exact deposition conditions of the Mo layer and the superconducting/normal metal interface. Because the individual contributions (strain, crystalline structure, contamination) are difficult to disentangle and control, reproducibility remains a challenge. Recently, we have demonstrated that low-energy ion-beam-assisted e-beam evaporation offers an alternative route to reliably produce high-quality Mo films without the use of substrate heating. The energy and momentum delivered by the ion beam provides an additional control knob to tune film properties such as resistivity and stress. In this report we describe modifications made to the commercial end-Hall ion source to avoid iron contamination allowing us to produce superconducting Mo films. We show that the ion beam is effective at enhancing the bilayer interface transparency and that bilayers can be further tuned towards reduced Tc and higher conductivity by vacuum annealing.
Cryogenic X-ray spectrometers are advantageous in the spectrum research for weak and diffusive X-ray source due to their high energy resolution, high detection efficiency, low noise level and non-dead-layer properties. Their energy resolution independent of the incident X-ray direction also makes them competitive in diffusion source detection. The requirements for X-ray spectrometers have heightened in recent years with the rapid development of large scientific facilities where X-ray detection is demanded, including beamline endstations in synchrotron and X-ray free electron laser facilities, accelerators, highly charged ion traps, X-ray space satellites, etc. Because of their excellent performances, cryogenic X-ray detectors are introduced into these facilities, typical examples of which are APS, NSLS, LCLS-II, Spring-8, SSNL, ATHENA, HUBS. In this paper, we review the cryogenic X-ray spectrometers, from the working principle and classification, system structure, major performance characteristics to the research status and trend in large scientific facilities in the world.