Magnetic resonance imaging (MRI) is among the most important medical imaging modalities. Coil arrays and receivers with high channel counts (16 and more) have to be deployed to obtain the image quality and acquisition speed required by modern clinical protocols. In this paper, we report the theoretical analysis, the system-level design, and the circuit implementation of the first receiver IC (RXIC) for clinical MRI fully integrated in a modern CMOS technology. The dual-channel RXIC sits directly on the sensor coil, thus eliminating any RF cable otherwise required to transport the information out of the magnetic field. The first stage LNA was implemented using a noise-canceling architecture providing a highly reflective input used to decouple the individual channels of the array. Digitization is performed directly on-chip at base-band by means of a delta-sigma modulator, allowing the subsequent optical transmission of data. The presented receiver, implemented in a $\mathbf {{130}\,{nm}}$ CMOS technology, is compatible with MRI scanners up to $\mathbf {{10.5}\,{T}}$. It reaches sub-$\mathbf {{1}\,{dB}}$ noise figure for $\mathbf {{3\!-\! 7}\,{T}}$ MRI units and features a dynamic range up to $\mathbf {{81.9}\,{dB}}$ at a power consumption below $\mathbf {{240}\,{mW}}$ per channel, with an area occupation of $\mathbf {{22}\,{mm^2}}$. Mounted on a small-sized printed circuit board (PCB), the receiver IC has been employed in a commercial MRI scanner to acquire in-vivo images matching the quality of traditional systems, demonstrating the first step toward multichannel wearable MRI array coils.
11 October 2018 in IEEE Transactions on Biomedical Engineering
The eGaIn coil on neoprene demonstrated in this paper presents a stretchable radio frequency receive coil for MRI. The coil with dimensions 86mm × 70mm is tuned to resonate at 128MHz for 3T MRI. We investigate the effect of stretching (up to 40% strain) and bending (50mm radius of curvature) of the coil on the coil's resistance and resonance frequency. Measurements and simulations show a decrease in resonance frequency of 2.5MHz per 10% strain. The higher resistivity of liquid metal compared to copper reduces the SNR of MRI scans by 2.1dB; therefore a trade-off between flexibility and performance remains. Nevertheless, we have successfully performed MRI scans with the liquid metal coil.