Laboratories with sensitive instruments need a low-vibration environment. It is a challenge to control the train-induced vibration impact on these instruments when a newly planned metro line is adjacent to a laboratory building. An alternative method of mitigating train-induced ground vibrations involves installing measures along the transmission path. Recent research has highlighted the potential of periodic pile barriers with specifically designed band gaps for controlling environmental vibrations. This study performed in-situ measurements of ambient vibrations inside and outside a laboratory containing various types of sensitive instruments and located adjacent to a newly designed metro line. The vibration transfer function of the laboratory was then obtained. To help design and optimize the band gaps of periodic piles, a novel band gap performance evaluation function was proposed. Finally, numerical analysis was conducted to validate the mitigation effect of the designed periodic piles. The results showed that the band gap performance evaluation function can be used to optimize the mitigation effect of periodic piles. The proposed periodic piles clearly attenuated vibrations between 52.4 and 74.3 Hz, especially those at 63 Hz. A comparison of general vibration criteria (VC) curves revealed that vibration attenuation of one level can be obtained by the designed periodic piles.
To predict and control the train-induced vibration in depot buildings, a case study of the depot of Tianjin Metro, Line 5, was conducted. The platform of the depot has been constructed and is in use, and the construction of over-track buildings has not been completed. Firstly, an in situ measurement was performed to obtain the train loads and validate the numerical model. Secondly, a finite element model of the track–soil–depot structure was established. The train was simplified as a series of two spring-mass models and the train load was simulated using the measured rail acceleration. The calculated results were validated by the measurement data. To predict the vibration responses of the over-track building to be built, a sub-system of the over-track building was added to the numerical model. Finally, the vibration control effect of vibration isolation bearings was discussed. The results indicate that vibrations exceeded guideline limits without mitigation measures in some rooms of the over-track building. The dominant frequency of the building floors is 31.5 Hz. Vibration isolation bearings effectively mitigated the vibrations, and the IL reached approximately 7–15 dB at about 31.5 Hz.
14 March 2022 in Journal of Computational and Nonlinear Dynamics
This article introduces a pendulum element to a 3-spring vibration isolator to achieve a high-static-low-dynamic (HSLD) stiffness or even quasi-zero stiffness (QZS) around the equilibrium position. The model is first established, the equilibrium point is derived and the optimal stiffness ratio of this novel system at the equilibrium position is also obtained. Numerical simulation is given and the harmonic balance method (HBM) is used to obtain time responses for analysis. Effects of different parameters on the isolation performance are studied and summarized. Approximation force and displacement transmissibility of the system are calculated to evaluate the isolation performance. Comparisons are made with those of an equivalent linear isolator and the typical 1 degree-of-freedom (DOF) QZS isolator. Results show that the novel vibration isolator performs better than existing isolators under selected parameters. The left bent backbone of the novel isolator demonstrates evident softening geometric nonlinearity. Therefore, it achieves a wider frequency range of isolation than the linear 1DOF isolator and typical 3-spring QZS isolator. Moreover, the transmissibility of the novel isolator is smaller at higher frequencies as the jump phenomenon occurs on the left.
Floating slab tracks (FSTs) are used to reduce the impact of vibration on precision instruments and historical relics along metro lines; however, ground vibration is universally amplified at the natural frequency of the tracks. In this study, a full-frequency control method that considers frequency matching for environmental vibrations, in combination with metro vibration sources and propagation paths, was developed based on the bandgap theory of the periodic structure. The effectiveness of this method was analysed by establishing a three-dimensional metro train–FST coupled model and a finite element analysis model of track bed–tunnel–soil–row piles. The results show that ground vibration can be reduced by approximately 3–5 dB at the natural frequency of the FST by adjusting the bandgap range of the periodic piles to 7–9 Hz, eliminating the adverse effect of vibration amplification at the natural frequency of the FSTs. The proposed control method shows good vibration control effects and can effectively minimise ground vibration in the full-frequency range.