Some of the recently released smartwatch products feature a single-lead electrocardiogram (ECG) recording capability. The reliability of obtaining 3-lead ECG with smartwatches is yet to be confirmed in a large study. This study aimed to assess the feasibility and reliability of smartwatch to obtain 3-lead ECG recordings, the classical Einthoven ECG leads I-III compared to standard ECG. To record lead I, the watch was worn on the left wrist and the right index finger was placed on the digital crown for 30 s. For lead II, the watch was placed on the lower abdomen and the right index finger was placed on the digital crown for 30 s. For lead III, the same process was repeated with the left index finger. Spearman correlation and Bland-Altman tests were used for data analysis. A total of 300 smartwatch ECG tracings were successfully obtained. ECG waves’ characteristics of all three leads obtained from the smartwatch had a similar duration, amplitude, and polarity compared to standard ECG. The results of this study suggested that the examined smartwatch (Apple Watch Series 4) could obtain 3-lead ECG tracings, including Einthoven leads I, II, and III by placing the smartwatch on the described positions.
The Apple Watch is capable of recording single-lead electrocardiograms (ECGs). To incorporate such devices in routine medical care, the reliability of such devices to obtain precordial leads needs to be validated. The purpose of this study was to assess the feasibility and reliability of a smartwatch (SW) to obtain precordial leads compared to standard ECGs. We included 100 participants (62 male, aged 62.8 ± 13.1 years) with sinus rhythm and recorded a standard 12-lead ECG and the precordial leads with the Apple Watch. The ECGs were quantitively compared. A total of 98 patients were able to record precordial leads without assistance. A strong correlation was observed between the amplitude of the standard and SW-ECGs’ waves, in terms of P waves, QRS-complexes, and T waves (all p-values < 0.01). A significant correlation was observed between the two methods regarding the duration of the ECG waves (all p-values < 0.01). Assessment of polarity showed a significant and a strong concordance between the ECGs’ waves in all six leads (91–100%, all p-values < 0.001). In conclusion, 98% of patients were able to record precordial leads using a SW without assistance. The SW is feasible and reliable for obtaining valid precordial-lead ECG recordings as a validated alternative to a standard ECG.
1 February 2022 in Indian Pacing and Electrophysiology Journal
This study evaluates the accuracy of iECGs in comparison to the gold standard ECG in adult patients with congenital heart disease and recommends the appropriate iECG derivation based on the patient's characteristics. In 106 adults (51 female, 55 male) with congenital heart disease, a gold standard 12-lead ECG was recorded, followed by three iECGs with the Apple Watch series 4, which correspond to Einthoven leads I, II, and III. Two experienced and independent cardiologists analyzed the time intervals, amplitudes, and polarities of the ECG parameters as well as the rhythm type and correlated the patient characteristics with the iECG parameters. The iECG parameters of all three iECG leads correlate strongly with those of the gold standard ECG, with exception of the P and T wave durations. We demonstrated that the informative value of the individual iECGs was independent of the patient's characteristics, in particular the heart axis, anatomy, and situs, even if the quality of the Einthoven III-like derivation was partially inadequate. The automatic rhythm analysis of the Apple Watch and the heart rhythm classification of a standard ECG analyzed manually by a cardiologist corresponded in 77%. iECG recordings of adults with congenital heart disease provide comparable results with Einthoven recordings I, II, and III of the 12-lead ECG and current data encourage the use of the Apple Watch not only in patients with structurally normal hearts but also in patients with congenital heart disease.
Background: Atrial fibrillation, the world's predominant cardiac arrhythmia, frequently emerges as a complication post-cardiac surgery, leading to serious outcomes like strokes, heart failures, and even death. Due to its often-silent nature, detecting it can be challenging. Smartwatches present a potential solution, offering screening that is more rigorous. Objective: This prospective observational study sought to assess the Withings Scanwatch's efficacy in identifying post-surgical atrial fibrillation. Methods: After cardiac surgery, patients received a Withings Scanwatch. Over a span of 24 hours, both the smartwatch's Photoplethysmography (PPG) sensor and standard telemetry kept track of any atrial fibrillation incidents. Results: At the end of the study, data from 260 patients was available for assessment. Atrial fibrillation was identified in 32 of these patients, either via telemetry or the smartwatch. Our data revealed a sensitivity of 69.0%, specificity of 98.7%, a PPV of 87.0%, and an NPV of 96.2%. Conclusions: This clinical study is the first to evaluate the PPG sensor of the Withings Scanwatch, and it shows that the Scanwatch has high accuracy in detecting atrial fibrillation. Thus, Scanwatch may support the conventional screening for atrial fibrillation, and potentially reducing complications and costs of atrial fibrillation. Because of lower than expected sensitivity this technology cannot replace conventional monitoring in postoperative patients.