Background: The concept of magnetic fluid hyperthermia is clinically evaluated after development of the whole body magnetic field applicator MFH® 300F and the magnetofluid MFL 082AS. This new system for localized thermotherapy is suitable either for hyperthermia or thermoablation. The magnetic fluid, composed of iron oxide nanoparticles dispersed in water, must be distributed in the tumour and is subsequently heated by exposing to an alternating magnetic field in the applicator. We performed a feasibility study with 22 patients suffering from heavily pretreated recurrences of different tumour entities, where hyperthermia in conjunction with irradiation and/or chemotherapy was an option. The potential to estimate (by post-implantation analyses) and to achieve (by improving the technique) a satisfactory temperature distribution was evaluated in dependency on the implantation technique. Material and methods: Three implantation methods were established: Infiltration under CT fluoroscopy (group A), TRUS (transrectal ultrasound) – guided implantation with X-fluoroscopy (group B) and intra-operative infiltration under visual control (group C). In group A and B the distribution of the nanoparticles can be planned prior to implantation on the basis of three-dimensional image datasets. The specific absorption rates (SAR in W/kg) can be derived from the particle distribution imaged via CT together with the actual H-field strength (in kA/m). The temperature distribution in the tumour region is calculated using the bioheat-transfer equation assessing a mean perfusion value, which is determined by matching calculated temperatures to direct (invasive or endoluminal) temperature measurements in reference points in or near the target region. Results: Instillation of the magnetic fluid and the thermotherapy treatments were tolerated without or with only moderate side effects, respectively. Using tolerable H-field-strengths of 3.0–6.0 kA/m in the pelvis, up to 7.5 kA/m in the thoracic and neck region and >10.0 kA/m for the head, we achieved SAR of 60–380 W/kg in the target leading to a 40°C heat-coverage of 86%. However, the coverage with ≥42°C is unsatisfactory at present (30% of the target volume in group A and only 0.2% in group B). Conclusion: Further improvement of the temperature distribution is required by refining the implantation techniques or simply by increasing the amount of nanofluid or elevation of the magnetic field strength. From the actual nanoparticle distribution and derived temperatures we can extrapolate, that already a moderate increase of the H-field by only 2 kA/m would significantly improve the 42°C coverage towards 100% (98%). This illustrates the great potential of the nanofluid-based heating technology.
Locoregional hyperthermia, i.e. increasing the tumor temperature to 40–45 °C using an external heating device, is a very effective radio and chemosensitizer, which significantly improves clinical outcome. There is a clear thermal dose-effect relation, but the pursued optimal thermal dose of 43 °C for 1 h can often not be realized due to treatment limiting hot spots in normal tissue. Modern heating devices have a large number of independent antennas, which provides flexible power steering to optimize tumor heating and minimize hot spots, but manual selection of optimal settings is difficult. Treatment planning is a very valuable tool to improve locoregional heating. This paper reviews the developments in treatment planning software for tissue segmentation, electromagnetic field calculations, thermal modeling and optimization techniques. Over the last decade, simulation tools have become more advanced. On-line use has become possible by implementing algorithms on the graphical processing unit, which allows real-time computations. The number of applications using treatment planning is increasing rapidly and moving on from retrospective analyses towards assisting prospective clinical treatment strategies. Some clinically relevant applications will be discussed.
We explored the non-thermal effects of radiofrequency (RF) electromagnetic fields and established a theoretical framework to elucidate their electrophysiological mechanisms. In experiments, we used a preclinical treatment device to treat the human colon cancer cell lines HT-29 and SW480 with either water bath heating (WB-HT) or 13.56 MHz RF hyperthermia (RF-HT) at 42 °C for 60 min and analyzed the proliferation and clonogenicity. We elaborated an electrical model for cell membranes and ion channels and estimated the resulting ion fluxes. The results showed that, for both cell lines, using RF-HT significantly reduced proliferation and clonogenicity compared to WB-HT. According to our model, the RF electric field component was rectified and smoothed in the direction of the channel, which resulted in a DC voltage of ~ 1 µV. This may induce ion fluxes that can potentially cause relevant disequilibrium of most ions. Therefore, RF-HT creates additional non-thermal effects in association with significant ion fluxes. Increasing the understanding of these effects can help improve cancer therapy.
1 January 2006 in International Journal of Hyperthermia
Background: Thermal treatments need verification of effectiveness. Invasive intra-tumoural thermometry was established as a standard method several years ago. However, in deep heating, invasive techniques have disadvantages. Therefore, alternatives have been suggested and are under development. Methods: In three phase II studies treating rectal cancer, cervical cancer and prostate cancer, this study replaced invasive (intra-tumoural) thermometry by tumour-related reference points or catheter sections in the rectum, vagina or urethra. Index temperatures and thermal dose parameters were determined. Two recent studies treated patients with recurrent rectal cancer and soft tissue sarcoma using non-invasive MR-thermometry employing the SIGMA-Eye applicator. The proton resonance frequency shift (PRFS) method was employed to generate MR-temperature distributions during the entire heat treatment in 10 min intervals (via phase differences). Fat correction (nulling specified regions in the fat tissue) was utilized to calibrate the method, in particular with respect to the B0-drift. Results: Statistically significant correlations were found between response (downstaging, WHO) and thermal parameters in rectal cancer (37 patients, rectum measurement, T90, cum min T90 ≥ 40.5°C) and cervical cancer (30 patients, vagina, mean temperature and equ min 43°C in a reference point). In prostate cancer (14 patients), a clear correlation was verified between long-term PSA control (≤1 ng ml−1) and urethral temperatures (T90, Tmax cum min T90 ≥ 40.5°C). The mean MR-temperature in the tumour at steady-state as well as the mean T90 were significantly correlated with response for recurrent rectal carcinoma regarding palliation and analgesia (15 patients) and with pathohistological regression rate in soft tissue sarcoma (nine patients). Conclusions: For tumours in the pelvis and in the lower extremities, invasive measurements can be replaced by minimally-invasive or non-invasive techniques, which provide equivalent or even more complete information. Extending the application of these surveillance methods to abdominal tumours or liver metastases is a challenge, but strongly desirable for clinical reasons.