In this work, we present a study of Aflatoxin M1 detection by photonic biosensors based on Si3N4 Asymmetric Mach–Zehnder Interferometer (aMZI) functionalized with antibodies fragments (Fab′). We measured a best volumetric sensitivity of 104 rad/RIU, leading to a Limit of Detection below 5 × 10−7 RIU. On sensors functionalized with Fab′, we performed specific and non-specific sensing measurements at various toxin concentrations. Reproducibility of the measurements and re-usability of the sensor were also investigated.
In this article, we designed and studied silicon oxynitride (SiON) microring-based photonic structures for biosensing applications. We designed waveguides, directional couplers, and racetrack resonators in order to measure refractive index changes smaller than 10−6 refractive index units (RIU). We tested various samples with different SiON refractive indexes as well as the waveguide dimensions for selecting the sensor with the best performance. Propagation losses and bending losses have been measured on test structures, along with a complete characterization of the resonator’s performances. Sensitivities and limit of detection (LOD) were also measured using glucose-water solutions and compared with expected results from simulations. Finally, we functionalized the resonator and performed sensing experiments with Aflatoxin M1 (AFM1). We were able to detect the binding of aflatoxin for concentrations as low as 12.5 nm. The results open up the path for designing cost-effective biosensors for a fast and reliable sensitive analysis of AFM1 in milk.
In this paper we report an innovative use of Poly(DiMethyl)Siloxane (PDMS) to design a microfluidic device that combines, for the first time, in one single reaction chamber, DNA purification from a complex biological sample (blood) without elution and PCR without surface passivation agents. This result is achieved by exploiting the spontaneous chemical structure and nanomorphology of the material after casting. The observed surface organization leads to spontaneous DNA adsorption. This property allows on-chip complete protocols of purification of complex biological samples to be performed directly, starting from cells lysis. Amplification by PCR is performed directly on the adsorbed DNA, avoiding the elution process that is normally required by DNA purification protocols. The use of one single microfluidic volume for both DNA purification and amplification dramatically simplifies the structure of microfluidic devices for DNA preparation. X-Ray Photoelectron Spectroscopy (XPS) was used to analyze the surface chemical composition. Atomic Force Microscopy (AFM) and Field Emission Scanning Electron Microscopy (FESEM) were employed to assess the morphological nanostructure of the PDMS-chips. A confocal fluorescence analysis was utilized to check DNA distribution inside the chip.