Polymer gel dosimeters offer a wide range of potential applications in the three-dimensional verification of complex dose distribution such as in intensity-modulated radiotherapy (IMRT). Until now, however, polymer gel dosimeters have not been widely used in the clinic. One of the reasons is that they are difficult to manufacture. As the polymerization in polymer gels is inhibited by oxygen, all free oxygen has to be removed from the gels. For several years this was achieved by bubbling nitrogen through the gel solutions and by filling the phantoms in a glove box that is perfused with nitrogen. Recently another gel formulation was proposed in which oxygen is bound in a metallo-organic complex thus removing the problem of oxygen inhibition. The proposed gel consists of methacrylic acid, gelatin, ascorbic acid, hydroquinone and copper(II)sulphate and is given the acronym MAGIC gel dosimeter. These gels are fabricated under normal atmospheric conditions and are therefore called 'normoxic' gel dosimeters. In this study, a chemical analysis on the MAGIC gel was performed. The composition of the gel was varied and its radiation response was evaluated. The role of different chemicals and the reaction kinetics are discussed. It was found that ascorbic acid alone was able to bind the oxygen and can thus be used as an anti-oxidant in a polymer gel dosimeter. It was also found that the anti-oxidants N-acetyl-cysteine and tetrakis(hydroxymethyl)phosphonium were effective in scavenging the oxygen. However, the rate of oxygen scavenging is dependent on the anti-oxidant and its concentration with tetrakis(hydroxymethyl)phosphonium being the most reactive anti-oxidants. Potentiometric oxygen measurements in solution provide an easy way to get a first impression on the rate of oxygen scavenging. It is shown that cupper(II)sulphate operates as a catalyst in the oxidation of ascorbic acid. We, therefore, propose some new normoxic gel formulations that have a less complicated chemical formulation than the MAGIC gel.
The employment of spectroscopically-resolved NMR techniques as analytical probes have previously been both prohibitively expensive and logistically challenging in view of the large sizes of high-field facilities. However, with recent advances in the miniaturisation of magnetic resonance technology, low-field, cryogen-free "benchtop" NMR instruments are seeing wider use. Indeed, these miniaturised spectrometers are utilised in areas ranging from food and agricultural analyses, through to human biofluid assays and disease monitoring. Therefore, it is both intrinsically timely and important to highlight current applications of this analytical strategy, and also provide an outlook for the future, where this approach may be applied to a wider range of analytical problems, both qualitatively and quantitatively.
A new method for the evaluation of radiotherapy 3D polymer gel dosimeters has been developed using ultrasound to assess the significant structural changes that occur following irradiation of the dosimeters. The ultrasonic parameters of acoustic speed of propagation, attenuation and transmitted signal intensity were measured as a function of absorbed radiation dose. The dose sensitivities for each parameter were determined as 1.8 x 10(-4) s m(-1) Gy(-1), 3.9 dB m(-1) Gy(-1) and 3.2 V(-1) Gy(-1) respectively. All parameters displayed a strong variation with absorbed dose that continued beyond absorbed doses of 15 Gy. The ultrasonic measurements demonstrated a significantly larger dynamic range in dose response curves than that achieved with previously published magnetic resonance imaging (MRI) dose response data. It is concluded that ultrasound shows great potential as a technique for the evaluation of polymer gel dosimeters.
Ultrasonic speed of propagation and attenuation were investigated as a function of absorbed radiation dose in PAG and MAGIC polymer gel dosimeters. Both PAG and MAGIC gel dosimeters displayed a dependence of ultrasonic parameters on absorbed dose with attenuation displaying significant changes in the dose range investigated. The ultrasonic attenuation dose sensitivity at 4 MHz in MAGIC gels was determined to be 4.7 +/- 0.3 dB m(-1) Gy(-1) and for PAG 3.9 +/- 0.3 dB m(-1) Gy(-1). Ultrasonic speed dose sensitivities were 0.178 +/- 0.006 m s(-1) Gy(-1) for MAGIC gel and -0.44 +/- 0.02 m s(-1) Gy(-1) for PAG. Density and compressional elastic modulus were investigated to explain the different sensitivities of ultrasonic speed to radiation for PAG and MAGIC gels. The different sensitivities were found to be due to differences in the compressional elastic modulus as a function of dose for the two formulations. To understand the physical phenomena underlying the increase in ultrasonic attenuation with dose, the viscoelastic properties of the gels were studied. Results suggest that at ultrasonic frequencies, attenuation in polymer gel dosimeters is primarily due to volume viscosity. It is concluded that ultrasonic attenuation significantly increases with absorbed dose. Also, the ultrasonic speed in polymer gel dosimeters is affected by changes in dosimeter elastic modulus that are likely to be a result of polymerization. It is suggested that ultrasound is a sufficiently sensitive technique for polymer gel dosimetry.