The traditional compensation model to explain the high resistivity properties of CdTe is based on the presence of a deep acceptor level of the cadmium vacancy in the middle of the band gap. A new compensation model based on a deep intrinsic donor level is presented. The compensation model is used together with an appropriate segregation model to calculate axial distributions of resistivity which are compared with spatially resolved resistivity measurements. The Te-antisite defect is discussed as a possible origin cause of this intrinsic defect, which is also supported by theoretical calculations.
2 December 2013 in IEEE Transactions on Nuclear Science
The spectroscopic performance of photon counting detectors is limited by the effects of charge sharing between neighboring pixels and the emission of characteristic X-rays. For these reasons, an event can be either missed or counted more than once. These effects become more and more of a concern when pixel pitches are reduced, and for the technology available so far, this meant that there would always be a trade-off between a high spatial and a high spectral resolution. In this work, we present first measurements obtained with the new Medipix3RX ASIC, which features a network of charge summing circuits establishing a communication between pixels which helps to mitigate these effects. Combined with cadmium telluride sensors, we show that this new technology is successful at improving a detector's spectroscopic capabilities even at pixel pitches as small as 55 μm. At this pitch, we measure an energy response function similar to that observed for a pixel pitch of 165 μm in the absence of a charge summing circuitry. This amounts to an effective reduction of the pixel area by at least one order of magnitude at a comparable energy response. Additionally, we present synchrotron measurements at high X-ray fluxes, where significant pulse pile-up occurs, and provide first experimental evidence for a net benefit when balancing spectroscopic performance and high flux tolerance in charge summing mode.
Spectroscopic x-ray imaging by means of photon counting detectors has received growing interest during the past years. Critical to the image quality of such devices is their pixel pitch and the sensor material employed. This paper describes the imaging properties of Medipix2 MXR multi-chip assemblies bump bonded to 1 mm thick CdTe sensors. Two systems were investigated with pixel pitches of 110 and 165 μm, which are in the order of the mean free path lengths of the characteristic x-rays produced in their sensors. Peak widths were found to be almost constant across the energy range of 10 to 60 keV, with values of 2.3 and 2.2 keV (FWHM) for the two pixel pitches. The average number of pixels responding to a single incoming photon are about 1.85 and 1.45 at 60 keV, amounting to detective quantum efficiencies of 0.77 and 0.84 at a spatial frequency of zero. Energy selective CT acquisitions are presented, and the two pixel pitches' abilities to discriminate between iodine and gadolinium contrast agents are examined. It is shown that the choice of the pixel pitch translates into a minimum contrast agent concentration for which material discrimination is still possible. We finally investigate saturation effects at high x-ray fluxes and conclude with the finding that higher maximum count rates come at the cost of a reduced energy resolution.