We report the quantum calibration of a magnetic force microscope (MFM) by measuring the two-dimensional magnetic stray-field distribution of the MFM tip using a single nitrogen vacancy (NV) center in diamond. From the measured stray-field distribution and the mechanical properties of the cantilever a calibration function is derived allowing to convert MFM images to quantum calibrated stray-field maps. This approach overcomes limitations of prior MFM calibration schemes and allows quantum calibrated nanoscale stray-field measurements in a field range inaccessible to scanning NV magnetometry. Quantum calibrated measurements of a stray-field reference sample allow its use as a transfer standard, opening the road towards fast and easily accessible quantum traceable calibrations of virtually any MFM.
Magnetic force microscopy (MFM) is a widespread technique for imaging magnetic structures with a resolution of some 10 nanometers. MFM can be calibrated to obtain quantitative (qMFM) spatially resolved magnetization data in units of A/m by determining the calibrated point spread function of the instrument, its instrument calibration function (ICF), from a measurement of a well-known reference sample. Beyond quantifying the MFM data, a deconvolution of the MFM image data with the ICF also corrects the smearing caused by the finite width of the MFM tip stray field distribution. However, the quality of the calibration depends critically on the calculability of the magnetization distribution of the reference sample. Here, we discuss a Ti/Pt/Co multilayer stack that shows a stripe domain pattern as a suitable reference material. A precise control of the fabrication process, combined with a characterization of the sample micromagnetic parameters, allows reliable calculation of the sample’s magnetic stray field, proven by a very good agreement between micromagnetic simulations and qMFM measurements. A calibrated qMFM measurement using the Ti/Pt/Co stack as a reference sample is shown and validated, and the application area for quantitative MFM measurements calibrated with the Ti/Pt/Co stack is discussed.
Magnetic skyrmions are topological spin structures that arise in chiral magnetic systems which exhibit broken inversion symmetry and high spin-orbit coupling resulting in a sizable Dzyaloshinskii-Moriya interaction. Understanding the local spin texture of skyrmions is a vital metrological step in the development of skyrmionic technologies required for novel logic or storage devices in addition to providing fundamental insight into the nanoscale chiral interactions inherent to these systems. Here, we propose that there exists a radially dependent stray field signature that emanates from magnetic skyrmions. We employ quantitative magnetic force microscopy to experimentally explore this stray field signature. To corroborate the experimental observations a semianalytical model is developed which is validated against micromagnetic simulations. This unique approach provides a route to understand the unique radially dependent field signature from skyrmions, which allows an understanding of the underlying local magnetization profile to be obtained. From a practical standpoint, our results provide a rapid approach to validate outputs from numerical or micromagnetic simulations. This approach could be employed to optimize the complex matrix of magnetic parameters required for fabricating and modeling skyrmionic systems, in turn accelerating the technology readiness level of skyrmionic based devices.
1 December 2025 in Review of Scientific Instruments
Magnetic force microscopy (MFM) allows the characterization of magnetic stray field distributions with high sensitivity and spatial resolution. Based on a suitable calibration procedure, MFM can also yield quantitative magnetic field values. This process typically involves measuring a reference sample to determine the distribution of the tip’s stray field or stray field gradient at the sample surface. This distribution is called the tip transfer function (TTF) and is derived through regularized deconvolution in Fourier space. The properties of the reference sample and the noise characteristics of the detection system significantly influence the derived TTF, thereby limiting its validity range. In a recent study, the tip stray field distribution, and hence the TTF, of an MFM tip was independently measured in real space using a nitrogen vacancy center as a quantum sensor, revealing considerable discrepancies with the reference-sample-based TTF. Here, we analyze the influence of the feature distribution of the reference sample and the MFM measurement parameters on the resulting TTF. We explain the observed differences between quantum-calibrated stray field distributions and the classical approach by attributing them to a loss of information due to missing or suppressed spectral components. Furthermore, we emphasize the importance of the spectral coverage of the TTF. Our findings indicate that for high-quality reconstruction of the stray field of a sample under test (SUT), it is more critical to ensure a strong overlap of frequency components between the reference sample and the SUT than to achieve an accurate real-space reconstruction of the tip stray field distribution.
24 September 2021 in Japanese Journal of Applied Physics
Pt( x Å)/Co(5 Å)/Pt(10 Å) trilayer films with x = 5, 8, and 10 Å were deposited on naturally oxidized Si substrate to investigate the Ti underlayer effect on perpendicular magnetic anisotropy. A small amount of Ti underlayer promoted the magnetic anisotropy to perpendicular direction for x = 8 and 10 Å and also affected the growth mode of the trilayer film. The Pt particles preferably grew in the neighborhood of the Ti cluster, giving rise to a relatively smooth layer with a fcc fiber texture structure. By contrast, the absence of a Ti underlayer led to a clustered, relatively rough, and randomly oriented nano-crystalline growth. Differences in the growth mode appeared as intensity loss in X-ray photoelectron spectroscopy spectra due to shadowing effects. Our results indicate that traceable amount of underlayer can change the magnetic anisotropy of the film by influencing the film growth.