Purpose: Modern facilities for actively scanned ion beam radiotherapy allow in principle the use of helium beams, which could present specific advantages, especially for pediatric tumors. In order to assess the potential use of these beams for radiotherapy, i.e., to create realistic treatment plans, the authors set up a dedicated 4He beam model, providing base data for their treatment planning system TRiP98, and they have reported that in this work together with its physical and biological validations.Methods: A semiempirical beam model for the physical depth dose deposition and the production of nuclear fragments was developed and introduced in TRiP98. For the biological effect calculations the last version of the local effect model was used. The model predictions were experimentally verified at the HIT facility. The primary beam attenuation and the characteristics of secondary charged particles at various depth in water were investigated using 4He ion beams of 200 MeV/u. The nuclear charge of secondary fragments was identified using a ΔE/E telescope. 3D absorbed dose distributions were measured with pin point ionization chambers and the biological dosimetry experiments were realized irradiating a Chinese hamster ovary cells stack arranged in an extended target.Results: The few experimental data available on basic physical processes are reproduced by their beam model. The experimental verification of absorbed dose distributions in extended target volumes yields an overall agreement, with a slight underestimation of the lateral spread. Cell survival along a 4 cm extended target is reproduced with remarkable accuracy.Conclusions: The authors presented a simple simulation model for therapeutical 4He beams which they introduced in TRiP98, and which is validated experimentally by means of physical and biological dosimetries. Thus, it is now possible to perform detailed treatment planning studies with 4He beams, either exclusively or in combination with other ion modalities.
The purpose of this work was to design and manufacture a 3D range-modulator for scanned particle therapy. The modulator is intended to create a highly conformal dose distribution with only one fixed energy, simultaneously reducing considerably the treatment time. As a proof of concept, a 3D range-modulator was developed for a spherical target volume with a diameter of 5 cm, placed at a depth of 25 cm in a water phantom. It consists of a large number of thin pins with a well-defined shape and different lengths to modulate the necessary shift of the Bragg peak. The 3D range-modulator was manufactured with a rapid prototyping technique. The FLUKA Monte Carlo package was used to simulate the modulating effect of the 3D range-modulator and the resulting dose distribution. For that purpose, a special user routine was implemented to handle its complex geometrical contour. Additionally, FLUKA was extended with the capability of intensity modulated scanning. To validate the simulation results, dose measurements were carried out at the Heidelberg Ion Beam Therapy Center with a 400.41 MeV/u 12 C beam. The high resolution dosimetric measurements show a good agreement between simulated and measured dose distributions. Irradiation of the monoenergetic raster plan took 3 s, which is approximately 20 times shorter than a comparable plan with 16 different energies. The combination of only one energy and a 3D range-modulator leads to a tremendous decrease in irradiation time. ‘Interplay effects’, typical for moving targets and pencil beam scanning, can be immensely reduced or disappear completely, making the delivery of a homogeneous dose to moving targets more reliable. Combining high dose conformity, very good homogeneity and extremely short irradiation times, the 3D range-modulator is considered to become a clinically applicable method for very fast treatment of lung tumours.
The radiation used in hadrontherapy treatments interacts with the patient body producing secondary particles, either neutral or charged, that can be used for dose and Bragg peak monitoring and to provide a fast feedback on the treatment plans. Recent results obtained from the authors on simplified setups (mono-energetic primary beams interacting with homogeneous tissue-like target) have already indicated the correlation that exists between the flux of these secondaries coming from the target (e.g. protons and photons) and the position of the primary beam Bragg peak. In this paper, the measurements of charged particle fluxes produced by the interaction of a 220 MeV/u carbon ion beam at GSI, Darmstadt, with a polymethyl methacrylate target are reported. The emission region of protons (p), deuterons (d) and tritons (t) has been characterized using a drift chamber while the particle time-of-flight, used to compute the kinetic energy spectra, was measured with a LYSO scintillator. The energy released in the LYSO crystal was used for particle identification purposes. The measurements were repeated with the setup at 60° and 90° with respect to the primary beam direction. The accuracy on the fragments emission profile reconstruction and its relationship with the Bragg peak position have been studied. Based on the acquired experimental evidence, a method to monitor the dose profile and the position of the Bragg peak inside the target is proposed.
Irradiation of moving targets using a scanned ion beam can cause clinically intolerable under- and overdosages within the target volume due to the interplay effect. Several motion mitigation techniques such as gating, beam tracking and rescanning are currently investigated to overcome this restriction. To enable detailed experimental studies of potential mitigation techniques a complex thorax phantom was developed. The phantom consists of an artificial thorax with ribs to introduce density changes. The contraction of the thorax can be controlled by a stepping motor. A robotic driven detector head positioned inside the thorax mimics e.g. a lung tumour. The detector head comprises 20 ionization chambers and 5 radiographic films for target dose measurements. The phantom's breathing as well as the 6D tumour motion (3D translation, 3D rotation) can be programmed independently and adjusted online. This flexibility allows studying the dosimetric effects of correlation mismatches between internal and external motions, irregular breathing, or baseline drifts to name a few. Commercial motion detection systems, e.g. VisionRT or Anzai belt, can be mounted as they would be mounted in a patient case. They are used to control the 4D treatment delivery and to generate data for 4D dose calculation. To evaluate the phantom's properties, measurements addressing reproducibility, stability, temporal behaviour and performance of dedicated breathing manoeuvres were performed. In addition, initial dosimetric tests for treatment with a scanned carbon beam are reported.
The roadmap for space exploration foresees longer journeys and further excursions outside low-Earth orbit as well as the establishment of permanent outposts on other celestial bodies, such as the Moon or Mars. The design of spacecraft and habitats depends heavily on the mission scenario and must consider the radiation protection properties of the structural components as well as dedicated shielding. In fact, short- and long-term effects caused by exposure to cosmic radiation are now considered among the main health risks of space travel. One of the current strategies is to find multifunctional materials that combine excellent mechanical properties with a high shielding effectiveness to minimize the overall load. In this work, the shielding effectiveness of a wide variety of single and multilayer materials of interest for different mission scenarios has been characterized. In the experimental campaign, reference and innovative materials, as well as simulants of Moon and Mars in situ resources, were irradia...
Measured cross sections for the production of the PET isotopes , and from carbon and oxygen targets induced by protons (40–220 ) and carbon ions (65–430 ) are presented. These data were obtained via activation measurements of irradiated graphite and beryllium oxide targets using a set of three scintillators coupled by a coincidence logic. The measured cross sections are relevant for the PET particle range verification method where accurate predictions of the emitter distribution produced by therapeutic beams in the patient tissue are required. The presented dataset is useful for validation and optimization of the nuclear reaction models within Monte Carlo transport codes. For protons the agreement of a radiation transport calculation using the measured cross sections with a thick target PET measurement is demonstrated.