This study aimed to evaluate the reliability of the RSQ Motion sensor and its validity against the Propriometer and electronic goniometer in measuring the active range of motion (ROM) of the shoulder. The study included 15 volunteers (mean age 24.73 ± 3.31) without any clinical symptoms with no history of trauma, disease, or surgery to the upper limb. Four movements were tested: flexion, abduction, external and internal rotation. Validation was assessed in the full range of active shoulder motion. Reliability was revised in full active ROM, a fixed angle of 90 degrees for flexion and abduction, and 45 degrees for internal and external rotation. Each participant was assessed three times: on the first day by both testers and on the second day only by one of the testers. Goniometer and RSQ Motion sensors showed moderate to excellent correlation for all tested movements (ICC 0.61–0.97, LOA < 23 degrees). Analysis of inter-rater reliability showed good to excellent agreement between both testers (ICC 0.74–0.97, LOA 13–35 degrees). Analysis of intra-rater reliability showed moderate to a good agreement (ICC 0.7–0.88, LOA 22–37 degrees). The shoulder internal and external rotation measurement with RSQ Motion sensors is valid and reliable. There is a high level of inter-rater and intra-rater reliability for the RSQ Motion sensors and Propriometer.
The invention of inertial measurement units allowed the construction of sensors suitable for human motion tracking that are more affordable than expensive optical motion capture systems, but there are a few factors influencing their accuracy, such as the calibration methods and the fusion algorithms used to translate sensor readings into angles. The main purpose of this study was to test the accuracy of a single RSQ Motion sensor in comparison to a highly precise industrial robot. The secondary objectives were to test how the type of sensor calibration affects its accuracy and whether the time and magnitude of the tested angle have an impact on the sensor’s accuracy. We performed sensor tests for nine repetitions of nine static angles made by the robot arm in eleven series. The chosen robot movements mimicked shoulder movements in a range of motion test (flexion, abduction, and rotation). The RSQ Motion sensor appeared to be very accurate, with a root-mean-square error below 0.15°. Furthermore, we found a moderate-to-strong correlation between the sensor error and the magnitude of the measured angle but only for the sensor calibrated with the gyroscope and accelerometer readings. Although the high accuracy of the RSQ Motion sensors was demonstrated in this paper, they require further study on human subjects and comparisons to the other devices known as the gold standards in orthopedics.
3 September 2016 in Springer Proceedings in Mathematics & Statistics