PurposeCollimator angle selection in stereotactic 3D conformal radiotherapy (3D-CRT) affects plan quality and delivery efficiency, especially in C-arm linear accelerators with multi-leaf collimators (MLC). This study evaluates whether heuristic collimator orientations, including zero-degree, computed optimal collimator angle (COCA), and combination strategies, complemented by Bayesian isocenter optimization, improve target coverage compared to expert-opinion planning in SBRT and fSRS.MethodsForty high-precision stereotactic brain and body cases were replanned for a linear accelerator with a 5-mm leaf-width MLC at isocenter. Collimator angles were varied across expert-defined beam arrangements, fixed 0°, COCA, and combined COCA strategies. Bayesian isocenter optimization was applied. Endpoints included MU per cGy, Conformity Index (CI_RTOG), Paddick Conformity Index (CIPaddick), Homogeneity Index (HI), Paddick Gradient Index (PGI), equivalent fall-off distance (EFOD), and geometric conformity (CIgeometric). Paired statistical tests were applied.ResultsCOCA-based strategies significantly reduced the average MU per cGy compared with expert planning, indicating improved delivery efficiency. The combined COCA and COCA+45° strategy achieved a 3.2% reduction (1.94 vs 2.00, p<0.001), indicating improved delivery efficiency. COCA also improved the CIRTOG by 4.6% (1.49 vs 1.56, p=0.003) and CIPaddick by 3.4% (0.66 vs 0.64, p=0.008) with a 1.6% increase in HI (0.70 vs 0.69, p≤0.009). Combined COCA strategies showed the largest efficiency gains. CIgeometric deteriorated for some strategies, such as COCA+45°, which increased by 4.1% (1.19 vs 1.15, p<0.00001). No significant changes were observed for PGI or EFOD.ConclusionHeuristic collimator angle optimization in SBRT/fSRS 3D-CRT improves delivery efficiency, conformity, and homogeneity, without affecting dose gradient metrics.