A novel exoskeleton robotic system was developed to assist stair climbing. This active demonstrator consists of a motor with a cable system, various sensors, and a control system with a power supply. The objective of this preliminary study is a biomechanical evaluation of the novel system to determine its effectiveness in use. For this purpose, three test persons were biomechanically investigated, who performed stair ascents and descents with and without the exoskeleton. Kinematics, kinetics, and muscle activity of the knee extensors were measured. The measured data were biomechanically simulated in order to evaluate the characteristics of joint angles, moments, and reaction forces. The results show that the new exoskeleton assists both the ascent and the descent according to the measured surface electromyography (sEMG) signals, as the knee extensors are relieved by an average of 19.3%. In addition, differences in the interaction between the test persons and the system were found. This could be due to a slightly different operation of the assisting force or to the different influence of the system on the kinematics of the users.
Climbing stairs can become a daily obstacle for elderly people, and an exoskeleton can assist here. However, the exoskeletons that are designed to assist stair climbing are actuated in different ways. To find a minimal actuation configuration, we identify the assist phases by evaluating the power deficit of 11 healthy but weak elderly people (72.4 ± 2.1 years; 69–76 years; 1.67 ± 0.10 m; 74.88 ± 14.54 kg) compared to 13 younger people (24.0 ± 1.8 years; 22–28 years; 1.74 ± 0.10 m; 70.85 ± 11.91 kg) in a biomechanical study and discuss moment characteristics. Three-dimensional kinematics and ground reaction forces were collected, and kinematics, kinetics, and power characteristics of each subject for ascent and descent were calculated using inverse dynamics. Significant differences for power between both groups were assessed with statistical parametric mapping method using dynamic time warping. During ascent, the largest significant power deficit of the elderly subjects occurs in the single stance phase (SSP) during pull-up in the knee joint. During descent, significant mean power deficits of 0.2 and 0.8 W/kg for the highest deficit occur in the ankle joint in the beginning of the SSP and also in the knee joint in the same phase. Therefore, an exoskeleton should address the power deficit for knee extension (ascent: 1.0 ± 0.9 W/kg; descent: 0.3 ± 0.2 W/kg) and could assist the ankle during ascent and descent by an additional plantar flexion moment of 0.2 Nm/kg each.