Fractures of the superior pubic ramus can be treated with screw insertion into the osseous fixation pathway (OFP) of the anterior column (AC). The entry point determines whether the screw exits the OFP prematurely. This can be harmful when it enters the hip joint or damages soft tissues inside the lesser pelvis. The exact entry point varies between patients and can be difficult to ascertain on fluoroscopy during surgery. The aim of this study was to determine variation in the location of the entry point. A retrospective single center study was performed at a level 1 trauma center in the Netherlands. Nineteen adult patients were included with an undisplaced fracture of the superior pubic ramus on computer tomography (CT)-scan. Virtual three-dimensional (3D) models of the pelvises were created. Multiple screws were placed per AC and the models were superimposed. A total of 157 screws were placed, of which 109 did not exit the OFP prematurely. A universally reproducible entry point could not be identified. A typical crescent shaped region of entry points did exist and was located more laterally in females when compared to males. Three-dimensional virtual surgery planning can be helpful to identify the ideal entry points in each case.
The conventional treatment for distal radius fractures typically involves immobilization of the injured extremity using a conventional forearm cast. These casts do cause all sorts of discomfort during wear and impose life-style restrictions on the wearer. Personalized 3D printed splints, designed using three-dimensional (3D) imaging systems, might overcome these problems. To obtain a patient specific splint, commercially available 3D camera systems are utilized to capture patient extremities, generating 3D models for splint design. This study investigates the feasibility of utilizing a new camera system (SPENTYS) to capture 3D surface scans of the forearm for the design of 3D printed splints. In a prospective observational cohort study involving 17 healthy participants, we conducted repeated 3D imaging using both the new (SPENTYS) and a reference system (3dMD) to assess intersystem accuracy and repeatability. The intersystem accuracy of the SPENTYS system was determined by comparison of the 3D surface scans with the reference system (3dMD). Comparison of consecutive images acquired per device determined the repeatability. Feasibility was measured with system usability score questionnaires distributed among professionals. The mean absolute difference between the two systems was 0.44 mm (SD:0.25). The mean absolute difference of the repeatability of the reference -and the SPENTYS system was respectively 0.40 mm (SD: 0.30) and 0.53 mm (SD: 0.25). Both repeatability and intersystem differences were within the self-reported 1 mm. The workflow was considered easy and effective, emphasizing the potential of this approach within a workflow to obtain patient specific splint.