Objectives: Measurement location and patient state can impact noninvasive liver assessment and change clinical staging in ultrasound examinations. Research into differences exists for Shear Wave Speed (SWS) and Attenuation Imaging (ATI), but not for Shear Wave Dispersion (SWD). The aim of this study is to assess the effect of breathing phase, liver lobe, and prandial state on SWS, SWD, and ATI ultrasound measurements. Methods: Two experienced examiners performed SWS, SWD, and ATI measurements in 20 healthy volunteers using a Canon Aplio i800 system. Measurements were taken in the recommended condition (right lobe, following expiration, fasting state), as well as (a) following inspiration, (b) in the left lobe, and (c) in a nonfasting state. Results: SWS and SWD measurements were strongly correlated (r = 0.805, p < 0.001). Mean SWS was 1.34 ± 0.13 m/s in the recommended measurement position and did not change significantly under any condition. Mean SWD was 10.81 ± 2.05 m/s/kHz in the standard condition and significantly increased to 12.18 ± 1.41 m/s/kHz in the left lobe. Individual SWD measurements in the left lobe also had the highest average coefficient of variation (19.68%). No significant differences were found for ATI. Conclusion: Breathing and prandial state did not significantly affect SWS, SWD, and ATI values. SWS and SWD measurements were strongly correlated. SWD measurements in the left lobe showed a higher individual measurement variability. Interobserver agreement was moderate to good.
The aim of the work described here was to investigate the relative contribution of confounding factors on liver shear wave speed (SWS) and shear wave dispersion slope (SWDS) measurements in three certified phantoms using a Canon Aplio clinical ultrasound scanner. A Canon Aplio i800 i-series ultrasound system (Canon Medical Systems Corporation, Otawara, Tochigi, Japan) with i8CX1 convex array (center frequency = 4 MHz) was used to examine dependencies caused by the depth, width and height of the acquisition box (AQB), the depth and size of the region of interest (ROI), the AQB angle and the pressure of the ultrasound probe on the surface of the phantom. Results revealed that depth is the most significant confounder in both SWS and SWDS measurements. AQB angle, height and width and ROI size exhibited minimal confounding effects on measurements. For SWS, the most consistent measurement depth is when the top of the AQB is placed between 2 and 4 cm, and the ROI is located between 3 and 7 cm deep. For SWDS, results indicate that measurement values significantly decrease with depth from the surface of the phantom until approximately 7 cm deep, and consequently no stable area of AQB placement or ROI depth exists. In contrast to SWS, the same ideal acquisition depth range cannot necessarily be applied to SWDS measurements because of a significant depth dependency.
There have been studies showing attenuation imaging (ATI) with ultrasound as an approach to diagnose liver diseases such as steatosis or cirrhosis. So far, this technique has only been used on a convex probe. The goal of the study was to investigate the feasibility of ATI measurements using the linear array on a canon Aplio i800 scanner on certified phantoms. Three certified liver tissue attenuation phantoms were measured in five different positions using a linear probe. The effects of positioning and depth were explored and compared. The values were compared to the certified expected value for each phantom as well as the different measurement values for each measurement position. The ATI measurements on phantoms showed significant effect for the different probe positions and region of interest (ROI) depths. Values taken in the center with the probe perpendicular to the phantom were closest to certified values. Median values at 2.5–4.5 cm depth for phantoms 1 and 2 and 0.5–2.5 cm for phantom 3 were comparable with certified values. Measurements taken at a depth greater than 6 cm in any position were the least representative of the certified values (p-value < 0.01) and had the widest range throughout the different sessions. ATI measurements can be performed with the linear probe in phantoms; however, careful consideration should be given to depth dependency, as it can significantly affect measurement values. Remaining measurements at various depths within the 0.5–6.0 cm range showed deviation from the certified values of approximately 25%.
Ultrasound attenuation imaging (ATI) is a non-invasive method for quantifying hepatic steatosis, offering advantages over the hepatorenal index (HRI). However, its reliability can be influenced by factors such as measurement depth, ROI size, and subcutaneous fat. This paper examines the impact of these confounders on ATI measurements and discusses diagnostic considerations. In this study, 33 healthy adults underwent liver ultrasound with ATI and HRI protocols. ATI measurements were taken at depths of 2–5 cm below the liver capsule using small and large ROIs. Two operators performed the measurements, and inter-operator variability was assessed. Subcutaneous fat thickness was measured to evaluate its influence on attenuation values. The ATI measurements showed a consistent decrease in attenuation coefficient values with increasing depth, approximately 0.05 dB/cm/MHz. Larger ROI sizes increased measurement variability due to greater anatomical heterogeneity. HRI values correlated weakly with ATI and were influenced by operator technique and subcutaneous fat, the latter accounting for roughly 2.5% of variability. ATI provides a quantitative assessment of hepatic steatosis compared to HRI, although its accuracy can vary depending on the depth and ROI selection. Standardised imaging protocols and AI tools may improve reproducibility and clinical utility, supporting advancements in ultrasound-based liver diagnostics for better patient care.
Background: and Methods: This paper presents a review of image quality in ultrasound imaging when using gel and liquid standoff pads, compared to imaging without a standoff pad. Imaging metrics assessed are lateral resolution, nearfield target intensity, and contrast to noise ratio of echogenicity targets. Measurements were made with a Canon Aplio i800 ultrasound system on a standardized ultrasound phantom (Computerized Imaging Reference Systems, Inc.), and custom-made agarose phantoms. Two cases are examined: when the ultrasound images are taken with the probe at different heights above the phantom (the normal clinical use-case for standoff pads); and when the ultrasound probe is artificially always placed at the same height above the phantom (allowing for direct image comparison). Results and Conclusion The main differences in the standard clinical use-case were observed in nearfield image intensity, with a mean difference of 22.4 ± 11.1 % between highest and lowest positions. Investigation of probes at the same height above the phantom shows that differences appear to be due primarily due to the height of the probe above the phantom, rather than the pads themselves. Elevational beamwidth was found to be a minimum of 2.2 mm at 25 mm axial depth, and adding a standoff pad can change vessel and cyst visualization due to the elevational beamwidth. This is a useful result to be aware of for clinicians using standoff pads on modern ultrasound systems.
Background/Objectives: Sarcopenia, an underdiagnosed musculoskeletal disorder, is a serious cause of disability, poor quality of life, and healthcare costs in an increasingly elderly population. This study aimed to examine an ultrasound (US)-based, inexpensive, simple, and reproducible alternative to magnetic resonance imaging (MRI) for assessing muscle quality. A study compared Dixon MR fat fraction with US attenuation imaging (ATI) and echo intensity (EI) in the rectus femoris (RF) and biceps brachii (BB). Methods: The US images were acquired from 34 participants who had previously received a whole-body MRI. The ATI measurements were carried out using a linear array on a Canon Aplio i800 scanner. The measurements of EI were assessed by manually tracing the cross-sectional border of the right RF and BB muscles. Corresponding T1-weighted Dixon VIBE-based fat and water images were required for the MRI fat fraction percentage (MR %FF) measurements. Results: Using Pearsons correlation coefficient, a good correlation was found between MR %FF and EI measurements. The results between operators’ measurements showed a strong correlation and were highly repeatable. Attenuation imaging revealed no correlation with MR %FF or EI. Conclusions: Echo intensity offers a low-cost, non-invasive, and widely accessible US-based imaging modality for screening patients at risk for sarcopenia. No correlation was found between the ATI and MR %FF or between the ATI and EI. Further adapted protocols and software adjustments are needed so that ATI has the potential to prove itself as an additional US-based method for assessing fat infiltration in muscles.