Miniature fixed-point blackbody (MFPB) can realize the on-orbit calibration of temperatures by tracing the miniature fixed-point. This ability can effectively improve the precision and life span of earth observation systems. Researchers have verified the feasibility of the MFPB in engineering through experiments. However, the temperature relationship between the miniature fixed-point and the blackbody is unclear at present, which means that the transfer relationship between radiation temperature of space infrared radiation standard source and international temperature scale cannot be established accurately. In this article, a temperature transfer relationship between miniature fixed-point and blackbody in MFPB is established, and the temperature transfer equation (TTE) is proposed to obtain the run-off point temperature of blackbody (RPTB). The effects of heating power and ambient temperature on the temperature transfer relationship between miniature fixed-point and blackbody are investigated numerically. Retrieval results show that a linear relationship exists between the RPTB and the heating power and ambient temperature in a single fixed-point temperature range. The TTE is obtained and explained theoretically. All experimental results of the gallium fixed-point are in good agreement with this equation. Furthermore, the RPTBs are accurately predicted at gallium-stannum and H₂O fixed-points by TTE. The average and maximum errors are 0.0044 and 0.0063 K, respectively. The current study is of considerable importance to the development of MFPB technology and improvement of the on-orbit infrared remote sensor calibration.
In this study, we realized the triple point (TP) temperature of sulfur hexafluoride (SF 6 ) at the National Institute of Metrology (NIM), China. To this end, we designed an iso-thermal and immersion type apparatus with an open cell with a height of 18 cm, diameter of 5 cm, and a capacity of approximately 275 ml. The cell was designed with four thermometer wells for the simultaneous temperature measurements with four standard platinum resistance thermometers (SPRTs). The cell is connected with two buffer tanks that allow a sufficient amount of SF 6 to surround the thermometer wells during the solidification and melting processes to ensure stable temperature plateaus with long durations. An SF 6 sample with a purity of 99.9997%, which is the highest purity currently available, was supplied and analyzed from the Korea Research Institute of Standards and Science. The SF 6 TP temperature was determined using four long-stem SPRTs. The melting plateaus obtained had a duration longer than 15 h, and the TP temperature of SF 6 was determined to be 223.556 03(54) K ( k = 1). These results are consistent with those of previous studies and further indicate that the SF 6 TP is a good candidate to replace mercury TP, which is one of the fixed points defined in the International Temperature Scale of 1990, for the calibration of SPRTs.
6 August 2011 in International Journal of Thermophysics