Frederic Juget

Dr. Frederic Juget

orcid

Research Focus & Metrics

Academic Career Length

14Years (2013-Present)

(Calculated once Scilit publication data is synced: current year − year of first publication + 1)

Representative Publications

View All Publications
14 March 2022 in EJNMMI Physics
Verified
Journal Article
Open access
PDF file
FJFrederic Juget
Frederic Juget
ZTZeynep Talip
Zeynep Talip
YNYoucef Nedjadi
Youcef Nedjadi
MDM. Teresa Durán
M. Teresa Durán
PGPascal V. Grundler
Pascal V. Grundler
CBClaude Bailat
Claude Bailat
Background: 161Tb draws an increasing interest in nuclear medicine for therapeutic applications. More than 99% of the emitted gamma and X-rays of 161Tb have an energy below 100 keV. Consequently, precise activity measurement of 161Tb becomes inaccurate with radionuclide dose calibrators when using inappropriate containers or calibration factors to account for the attenuation of this low energy radiation. To evaluate the ionization chamber response, the sample activity must be well known. This can be performed using standards traceable to the Système International de Référence, which is briefly described as well as the method to standardize the radionuclides. Methods: In this study, the response of an ionization chamber using different container types and volumes was assessed using 161Tb. The containers were filled with a standardized activity solution of 161Tb and measured with a dedicated ionization chamber, providing an accurate response. The results were compared with standardized solutions of high-energy gamma-emitting radionuclides such as 137Cs, 60Co, 133Ba and 57Co. Results: For the glass vial type with an irregular glass thickness, the 161Tb measurements gave a deviation of 4.5% between two vials of the same type. The other glass vial types have a much more regular thickness and no discrepancy was observed in the response of the ionization chamber for these type of vials. Measurements with a plastic Eppendorf tube showed stable response, with greater sensitivity than the glass vials. Conclusion: Ionization chamber measurements for low-energy gamma emitters (< 100 keV), show deviation depending on the container type used. Therefore, a careful selection of the container type must be done for activity assessment of 161Tb using radionuclide dose calibrators. In conclusion, it was highlighted that appropriate calibration factors must be used for each container geometry when measuring 161Tb and, more generally, for low-energy gamma emitters.

1 December 2016 in Applied Radiation and Isotopes
Verified
Journal Article
Closed access
FJFrederic Juget
Frederic Juget
YNYoucef Nedjadi
Youcef Nedjadi
TBThierry Buchillier
Thierry Buchillier
FBFrançois Bochud
François Bochud
CBClaude Bailat
Claude Bailat

31 January 2023 in Molecules
Verified
Journal Article
Open access
PDF file
FJFrederic Juget
Frederic Juget
MRMaria Teresa Duran Duran Ramiro
Maria Teresa Duran Duran Ramiro
YNYoucef Nedjadi
Youcef Nedjadi
ZTZeynep Talip
Zeynep Talip
PGPascal V. Grundler
Pascal V. Grundler
NMNick Van der Meulen
Nick Van der Meulen
44Sc is a promising radionuclide for positron emission tomography (PET) in nuclear medicine. As a part of the implementation of a production site for 44Sc, precise knowledge of the activity of the product is necessary. At the Paul Scherrer Institute (PSI) and the University of Bern (UniBE), 44Sc is produced by enriched 44CaO-target irradiation with a cyclotron. The two sites use different techniques for activity measurement, namely a dose calibrator at the PSI and a gamma-ray spectrometry system at UniBE and PSI. In this work, the 44Sc was produced at the PSI, and samples of the product were prepared in dedicated containers for onsite measurements at PSI, UniBE, and the Institute of Radiation Physics (IRA) in Lausanne for precise activity measurement using primary techniques and for the calibration of the reference ionization chambers. An accuracy of 1% was obtained for the activity measurement, allowing for a precise calibration of the dose calibrator and gamma-ray spectrometry of the two production sites. Each production site now has the capability of measuring 44Sc activity with an accuracy of 2%.

1 August 2021 in Applied Radiation and Isotopes
Verified
Journal Article
Open access
FJFrederic Juget
Frederic Juget
ZTZeynep Talip
Zeynep Talip
TBThierry Buchillier
Thierry Buchillier
MDM. Teresa Durán
M. Teresa Durán
YNYoucef Nedjadi
Youcef Nedjadi
CBClaude Bailat
Claude Bailat
In this study, the gamma and X-ray emission intensities of 161Tb were determined using a high-purity germanium spectrometer. The samples used were previously standardised by coincidence counting and Triple to Double Coincidence Ratio (TDCR) methods. A total of 28 gamma-rays and 4 X-rays were measured and compared with previous measurements performed more than 30 years ago. Most of the lines are in agreement, while large discrepancies are observed for 5 lines. The uncertainties have been dramatically decreased with respect to previous measurements giving a better knowledge of the 161 Tb day.

1 April 2018 in Applied Radiation and Isotopes
Verified
Journal Article
Closed access
FJFrederic Juget
Frederic Juget
YNYoucef Nedjadi
Youcef Nedjadi
TBThierry Buchillier
Thierry Buchillier
TDTeresa Durán
Teresa Durán
FBFrançois Bochud
François Bochud
CBClaude Bailat
Claude Bailat

19 November 2013 in Applied Radiation and Isotopes
Verified
Journal Article
Closed access
FJFrederic Juget
Frederic Juget
CBClaude Bailat
Claude Bailat
FBFrançois Bochud
François Bochud